Angel
®
cPRP System
Operator’s Manual
1-4
Overview
1. Open the shipping carton and remove the Angel
®
Concentrated Platelet
Rich Plasma (cPRP) System and all ancillary components from the shipping
carton. Install the device in an operating environment between 10 and 30
degrees C (50-86 degrees F) and 10%-75% relative humidity.
2. Place the Angel
®
Concentrated Platelet Rich Plasma (cPRP) System on a
solid, flat work surface.
3. Install the power cord supplied with the equipment into the Angel
®
Concentrated Platelet Rich Plasma (cPRP) System. Plug the power into an
available power outlet
PRECAUTION
There are no user-serviceable parts inside this device . To avoid
the risk of electrical shock, do not remove the cover . Refer all
servicing to qualified service personnel.
Unpacking/Assembly
1. Inspect the exterior of the shipping container and verify that no visual
damage is present.
A. If damage is present, notify Cytomedix, Inc. Customer Service Response
Center at 877-865-9927 to report the damage.
B. If no damage is present, proceed with the installation.
Visual Inspection
Power-up the Angel
®
Concentrated Platelet Rich Plasma (cPRP) System by moving
the power switch located at the back of the machine to the on position. On power-
up, the Angel
®
Concentrated Platelet Rich Plasma (cPRP) System will perform an
automatic self-test. At this time, the valve assembly will also calibrate and reposition
itself. Successful completion of that self-test will be evident by visual confirmation of
the following start-up screen:
Operational Checks
Figure 1-4 Load Screen
No other checks or tests are required as part of this installation procedure.