30
9
Settings
16.
NOTE:
‘tn-1’ is used to adjust the temperature of chambers 5 and 10
‘tn-2’ is used to adjust the temperature of chambers 4 and 9
‘tn-3’ is used to adjust the temperature of chambers 3 and 8
‘tn-4’ is used to adjust the temperature of chambers 2 and 7
‘tn-5’ is used to adjust the temperature of chambers 1 and 6
‘tn-6’ is the Dry Bath Chamber
17.
NOTE:
Adjust according to a high precision measurement performed in a dish located in the
chamber
18.
NOTE:
The INT.T (Integral Time) option is for changing the base value of the PID controller.
From the factory, it is set to a closely calculated value specific for the model
19.
NOTE:
There are no calendar functions in the timer so all days include Saturday
and Sunday
20.
NOTE:
For optimal air purification reasons, we recommend the UV light set is to ON when the
incubator is being used
21.
NOTE:
Inside pressure inside the incubator is always lower than that shown on the outside
regulator of the gas bottle. This is because there is a built-in pressure regulator inside the
incubator
22.
NOTE:
The decontamination program heats up the whole heated area (zones 1 to 5) up to
60°C for approximately one hour in order to decontaminate the heated surfaces. Chamber
11 (Dry Bath) is not an incubation chamber, and will not be decontaminated, but it will still be
heated above the setpoint
23.
NOTE:
The decontamination system is equipped with a timer that automatically stops the
decontamination process when the desired temperature has been reached, and it brings the
system back to normal status
24.
NOTE:
While the decontamination process is active all gas regulations are deactivated. An
audible alarm will warn the user about rising temperatures. The actual temperature and
decontamination time left will be shown on the display. The system will also warn the user
about hot surface temperatures
25.
NOTE:
After the decontamination process the incubator will cool down slowly. Do not use the
device until the temperature and gas levels are back to normal. We recommend testing the
temperature and gas concentrations before use, to ensure correct levels
26.
NOTE:
This function sets all parameters back to factory default settings
27.
NOTE:
The firmware installed on your G185 incubator can be updated. Ensure your incubator
runs with the latest firmware. Your distributor or local representative is informed of all updates
Front Chamber Keys
On the sloping front of the G185 incubator there are 10 buttons. Each button corresponds to one of
the chambers.
Pressing a key will toggle the light in the key on and off. This light can be used as an indication that the
chamber is in use and should not be opened. Turning the light on/off has no effect on the temperature
or gas control of the chamber.
NOTE:
Never leave a chamber lid open for unnecessary periods of time. Each opening should be
Button No.1 for No.1 chamber