20
8
Basic Operation
Section 8 - Basic Operation
CAUTION:
Do not use the incubator if the alarm system has issued a failure message
and the cause of the failure has not been corrected
It is important that the appropriate dish inserts are selected for the culture dishes used (Falcon, Nunc,
Vitrolife) to ensure there is direct contact between the dish and the heated surface.
Dishes which present no opportunity for an air gap between dish and heated surface can be placed
directly on the heated surface with no need for a dish insert.
The use of dish inserts requires temperature calibration with the dish insert in place (see “Dish Inserts”
on page 12)
1.
Place the dish inserts in the chambers and close the lids
2.
Wait 30 minutes for the dish inserts to heat up
3.
Open the chamber lid
4.
Place the culture dish containing gametes or embryos on the dish insert ensuring it is placed
securely in the correct milled grooves
5.
Close the lid
6.
Note the patient ID as per “Writing pads for chamber lids and penholder” on page 32
User Interface
The following will explain functions associated with the keyboard and menu items.
Main keys and their purposes.
Key
Meaning
Alarm button
On/Standby
Turns the unit ON or OFF (standby mode)