Settings
WARNING:
The embryos stored in each chamber should always be identified
Magnetic writing pads can be attached to each chamber lid for
patient details or chamber content. Only write with a suitable
nontoxic pen that allows text to be removed afterwards. Steel-
bodied pens can be stuck to the magnetic pad.
Writing Pads for Chamber Lids and Penholder
NOTE:
If a current source is attached to the connector, the maximum current rating is 1.0 Amp.
NOTE:
If a voltage source is attached then the limitation is between 0 – 24V AC or DC.
If there is no alarm, the switch will be in
‘ON’ position
Whenever the G185 incubator goes into
alarm mode the switch will become an
‘open circuit’. This means no current will
run through the system.
NOTE:
Whenever the incubator power
cord is removed from the power outlet
in the wall it will automatically indicate
an alarm. This is an extra safety feature
added to alert personnel of power cuts.