31
9
Settings
CAUTION:
Do not block the gas inlet/outlet in the middle of each chamber as this will
cause incorrect gas concentration in the chamber
Gas Inlet/Outlet
minimized to less than 10 seconds under normal use, otherwise it may result in an alarm.
The base of each chamber is a warmed area. The gas intake and outlet located in the middle of the
the chamber must be kept free at all times.
Timer
The G185 contains a timer that can turn on the unit at a specific time. The timer is set by activating the
On time and turning on the timer. The On time is set in the menu point Start Set (St.St). The hours are
counted up using the down key and the minutes are counted up using the up key. The timer is turned
On by setting the “heat” point in the menu to On.
The timer can be set to repeat every day by activating the Automatic Start (A-St) to On in the menu.
NOTE:
If the Heat is set to Off when the timer function is set, nothing will be switched on when the
timer engages.
NOTE:
Timer functions should only be used with short-term culturing and are not applicable for
overnight culturing.
Clock
The clock time can be set by activating the time set (ti.St) variable. To do this, use the down key to
count the hours up and the up key to count the minutes up. The clock is used with the timer and can
be displayed when the unit is off; this is done by setting ‘Hour’ in the menu to On.
Alarm Switch for External Monitoring System
The G185 incubator is equipped with a 3.5mm mini-jack connector (on the rear panel) which allows
the incubator to be connected to an external monitoring system.
If an alarm is activated, the switch is deactivated indicating that the unit needs inspection by a user.
The connector can be attached to a voltage source or to a current source.
Mini-jack port