Intended Use
12
4
Ensure the culture dishes are placed securely in the correct milled grooves of the Dish Inserts.
Product Description
The G185 is composed of 10 separate culture chambers (2) and one larger warming chamber (3).
The 10 culture chambers are exposed to both heat and gas mixtures while the Dry Bath Chamber is
exposed only to heat.
The gas is constantly cleaned for particles, volatile organic compounds and bacteria by circulation
through a HEPA and VOC filter and through a 254nm UV chamber.
The CO
2
concentration is monitored by using a CO
2
sensor and the O
2
level is monitored by an O
2
sensor placed in the gas mixing chamber (built inside the incubator).
Dish Inserts
The chambers should only be fitted with special Dish Inserts (1), that allow safe placement of standard
culture dishes (Falcon, Nunc, Vitrolife).