CAUTION
•
DO NOT
use the incubator if ShockWatch or TipNTell has been triggered or if the package is
damaged
•
Read and understand the manual completely before use. Keep the manual close to the unit
•
Never use or handle this unit in ways other than specified in this manual. Your safety may be at
risk and the unit may get damaged
•
Never try to move the unit without consulting Technical Support
•
Never use the unit if the alarm system of the device has issued a failure message and the cause
of the failure has not been identified
•
Protect the power cord from being damaged or being restricted in any way. Unplug the power
cord from the wall socket or at the rear of the instrument to disconnect the mains supply
•
Make sure that CO
2
and N
2
gas supply pressures are not above 1.0 bar and not below 0.5 bar
•
Always keep the red caps on unused gas inlets and the sample port at the back of the unit
•
Never use the unit without an original K-Systems H13 filter
•
DO NOT
expose the filter to liquids. Change filters that have been exposed to liquids
•
DO NOT
leave lids open for more than 10 seconds
•
DO NOT
use the unit at ambient temperatures exceeding 30°C. Ambient temperature above
30°C will compromise the incubation process. The relative humidity must not exceed 75% (non-
condensing)