50
13
Service
For the reliable and safe operation of this incubator, it is strongly recommended that inspections and
services are performed as stated in the Service Plan below. Failing to follow this plan may cause the
unit to stop performing as intended and cause damage to embryos etc. kept inside the incubator.
Service Plan
Conducted by
User
Authorised Service Representative
Component name
Every 3
months
Every
year
Every 3
years
Every 6
years
Replace H13 Filter Capsule
X
Replace UV Light Bulb
X
Replace HEPA Inline Filter for CO
2
gas
X
Replace HEPA Inline Filter for N
2
gas
X
Replace O
2
sensor*
X
Temperature and gas calibration
X
Ventilaton filter (optional)
X
Cooling fan
X
Internal blower
X
Replace CO
2
sensor*
X
*Gas calibration should be performed after replacing O
2
and CO
2
sensors.
WARNING:
DO NOT disassemble or modify any part of the G185
Section 13 - Service