14
ITEM NO.
NAME
FUNCTION
1
Ground Check Switch
Turn ON for use with three-wire (grounded) power supply cords.
Turn OFF for use with two-wire power supply cords.
(When ON, conducts Ground Test between the chassis and the grounding pin of the line cord of the
equipment being tested. When OFF, Ground Test is bypassed for testing of double-insulated equipment
and other types of equipment without a grounding pin in the line cord.)
2
Directions
Provides directions for tester operation to test personnel.
3
Ramp Time Adjustment
Used to adjust the amount of time used to increase the high voltage from zero volts to the required level.
See "Adjustment of the High Voltage Ramp Time" for this procedure.
4
Test Time Adjustment
Used to adjust high voltage test duration. See "Adjustment of the High Voltage Test Duration" section
for this procedure.
5
Leakage Limit Adjustment
Used to adjust the trip level for the excessive leakage current test. See "Adjustment of the Leakage
Current Level" section for this procedure.
6
Voltage Adjust
Used to adjust the High Voltage output. See "Adjustment of the High Voltage Level" section for this
procedure.
7
Calibration Resistor
(Optional)
8
Appliance Inlet / Fuse holder /
Power Switch
Use supplied cord set to connect tester to appropriate source of supply. Replace line fuse. Turn Tester
ON/OFF.
9
Fuse replacement warning /
Rating of supply
Specifies replacement fuse and required supply voltage.
Table 3. Controls, Indicators, Connectors - Model HT-3000P Rear Panel