background image

 

27 

B)      Press 

Test

  to  enter  the 

Calibration  Mode

  and  create  new  calibration  settings.    (Be 

sure you want to enter the 

Calibration Mode

 as this will change the laboratory number so 

it will show the calibration was not performed by Compliance West USA). 

6. 

Once  the 

Calibration  Mode

  has  been  entered,  the 

Reset

  button  toggles  between  the  5 

calibration menus: Volt, V1, V2, curr, and bars.  Make sure "V1" is displayed on the meter. 

7. 

Hook  up  a  calibrated  1000:1  voltage  probe  and  a  calibrated  DMM  between  the  Device 
receptacle and the Return jack on the front panel. 

8. 

Use  the  AC/DC  selector  switch  on  the  front  panel  to  choose  which  voltage  will  be 
calibrated. 

9. 

Turn the Voltage Adjust on the rear panel to minimum (counterclockwise).  Press the 

Test

 

button  and  a  voltage  number  will  be  displayed  on  the  front  panel  meter  (1500  for  AC  or 
2500 for DC).  Be careful as the HT-3000P will be putting out voltage at this point. 

10.  Turn the Voltage Adjust on the rear panel clockwise until the output on the DMM equals 

the appropriate voltage (1500 volts AC or 2500 volts DC) and press the Test button.  The 
display will now display "V1" again. 

11.  Press the 

Reset

 button until "V2" is displayed on the front panel meter. 

12.  Turn the Voltage Adjust on the rear panel to minimum (counterclockwise).  Press the 

Test

 

button  and  500  will  be  displayed  on  the  front  panel  meter  (for  either  AC  and  DC).    Be 
careful as the HT-3000P will be putting out voltage at this point. 

13.  Turn the Voltage Adjust on the rear panel clockwise until the output on the DMM equals 

the appropriate voltage (500 volts AC or DC) and press the Test button.  The display will 
now display "V2" again.  The voltage is now be calibrated. 

14.  Turn Off the HT-3000P tester. 
15.  The  voltage  can  be  verified  by  turning  on  the  HT-3000P,  running  a  test,  and  using  the 

1000:1 voltage probe to check the meter at different voltages.  If the voltages are not within 
+/-  2% at each voltage, the voltage calibration must be performed again starting at step 1 
above. 

Current Calibration Adjustment 

1. 

You must be in 

Calibration Mode

 as in step 3 under the Voltage Calibration Adjustment 

above. 

2. 

Use the AC/DC selector switch on the front panel to choose AC. 

3. 

Turn the Voltage Adjust on the rear panel to minimum (counterclockwise).  

4. 

Connect the output of the HT-3000P through a 60K ohm, 100 watt high voltage resistor in 
series with a calibrated DMM capable of measuring 10.0mA AC and returning to the 
Return jack on the front panel of the HT-3000P. 

5. 

In the calibration mode, hit the Reset button until "curr" is displayed on the front panel. 

6. 

Once "curr" is displayed, press the Test button once and 10.0 will be displayed. 

7. 

Slowly increase the variac watching the current flowing on the DMM.  Adjust the variac 
until 10.0mA AC is displayed on the DMM and then press the Test button. 

8. 

The display will show "hold". 

9. 

Wait until the display returns to "curr".  At this point the AC leakage current is calibrated. 

10. 

Turn the Voltage Adjust on the rear panel to minimum (counterclockwise). 

11. 

Use the AC/DC selector switch on the front panel to choose DC.  

12. 

Connect the output of the HT-3000P through a 240K ohm, 100 watt high voltage resistor in 
series with a calibrated DMM capable of measuring 5.0mA DC and returning to the Return 
jack on the front panel of the HT-2800P. 

13. 

In the calibration mode, hit the Reset button until "curr" is displayed on the front panel. 

14. 

Once "curr" is displayed, press the Test button once and 5.0 will be displayed. 

15. 

Slowly increase the variac watching the current flowing on the DMM.  Adjust the variac 
until 5.0mA DC is displayed on the DMM and then press the Test button. 

16. 

The display will show "hold". 

17. 

Wait until the display returns to "curr".  At this point the DC leakage current is calibrated. 

Summary of Contents for HT-3000P

Page 1: ...HT 3000P Dielectric Withstand Tester 0 2000 Volts AC Output 0 2800 Volts DC Output Instruction Manual COMPLIANCE WESTUSA...

Page 2: ......

Page 3: ...logic controlled circuit design and ergonomic front panel and represents the latest in high voltage production line testing To fully appreciate all the features of your new meter we suggest that you t...

Page 4: ...2...

Page 5: ...and Rear Panel Features 11 Initial Checkout Procedure 16 Setting up the HT 3000P for Production Line Testing 17 Factory Settings 17 Display of Leakage Limit and Duration settings 17 Adjustment of the...

Page 6: ...ndamentals clearly and be aware that high voltage is adept and creative at completing a path to ground Instructions should include a warning against any metal jewelry Operators should not allow others...

Page 7: ...tion between the operator and the test being conducted The HT 3000P is designed to bleed the high voltage away after the test has concluded In order to ensure that any voltage present in the equipment...

Page 8: ...continuously during the test When testing with the Ground Check switch in the Off position the operator must ensure that the Test Return Lead is properly connected to exposed dead metal of the equipme...

Page 9: ...s is due to the AC voltage charging all capacitors connected between primary voltage and ground If the overall value of these capacitors cause a leakage current of more than 20 mA to flow at the desir...

Page 10: ...ntil the voltage has discharged to a safe level to protect the operator and also to ensure that there is no energy stored in the tested equipment Testing Equipment with non Standard Plugs or Pigtail O...

Page 11: ...ff by a resistor bank in the HT 3000P The voltage discharge progress is shown by the front panel meter Failure modes are shown by the front panel LED s for quick troubleshooting Convenience and testin...

Page 12: ...available by removing adjustment shafts ENVIRONMENTAL Operating Temperature 15 40 C Relative Humidity Range 0 90 non condensing GENERAL Input power requirements See rear panel for input voltage and f...

Page 13: ...ock absorbing material on all sides of the container Remove the tester from its container and place it on a test bench AC Line Voltage Requirements AC line voltage requirements for your Tester are not...

Page 14: ...11 Figure 2 Controls Indicators Connectors Model HT 3000P Front Panel...

Page 15: ...d the test will be terminated 5 Full Voltage LED If full voltage is successfully reached the Full Voltage LED will light and the high voltage duration timer starts 6 Hipot Fail LED Indicates failure o...

Page 16: ...13 Figure 3 Controls Indicators Connectors Model HT 3000P Rear Panel...

Page 17: ...ent of the High Voltage Ramp Time for this procedure 4 Test Time Adjustment Used to adjust high voltage test duration See Adjustment of the High Voltage Test Duration section for this procedure 5 Leak...

Page 18: ...15...

Page 19: ...osition 8 Verify that all test leads are disconnected from the HT 3000P 9 Push the red RESET button then the yellow TEST button 10 The tester should sound the buzzer The red Ground Open and red RESET...

Page 20: ...may light and in some instances the Ground OK LED may extinguish Passage of these tests indicates that the HT 3000P is functioning properly and that it is safe to use If the results of the performanc...

Page 21: ...ons Use this procedure if adjustment to a different ramp time within the range of 1 5 seconds is required 1 Defeat the Ground Check by turning the rear panel switch to the Off position 2 Connect the H...

Page 22: ...the rear panel turn the Test Time potentiometer As soon as the potentiometer starts turning the meter will display d and the value can be set in 1 second increments from 1 to 60 seconds 4 If tamperpro...

Page 23: ...e power switch to the I or ON position Push the RESET button The yellow TEST indicator should light indicating that the HT 3000P is ready to test 2 Connect the HTT 1 to the HT 3000P tester using the c...

Page 24: ...f time set in the Test Duration procedure If a breakdown of the insulation system between primary voltage and ground is detected the red Hipot Fail LED will light the buzzer will sound and the test wi...

Page 25: ...ssistance Phone 800 748 6224 Technical Assistance is available from Compliance West USA between the hours of 8 30 AM and 4 30 PM Pacific Time Compliance West USA 2120 Jimmy Durante Blvd Suite 118 Del...

Page 26: ...23...

Page 27: ...ad with a banana plug on one end are required to conduct the performance test The test equipment required for the calibration procedure is a DMM able to read true rms 0 2000 Vac 1 0 2800 Vdc 1 and a 3...

Page 28: ...the front panel and case with a mild solution of detergent and a damp sponge Clean dust from the PWB with clean dry low pressure 20 psi Calibration Procedure The Calibration Procedure should be perfo...

Page 29: ...r clip end to the 3 ohm resistor 3 Connect the banana plug end of the other lead to the round grounding jack of the DEVICE receptacle and the alligator clip end to the 3 ohm resistor 4 Turn the Ground...

Page 30: ...a test and using the 1000 1 voltage probe to check the meter at different voltages If the voltages are not within 2 at each voltage the voltage calibration must be performed again starting at step 1 a...