19
3.
Connect the HT-3000P to a correctly rated source of supply and turn the power switch to
the I or ON position. Push the
RESET
button. The yellow
TEST
indicator should light,
indicating that the HT-3000P is ready to test.
4.
Use the AC/DC button to select the type of output voltage.
5.
Make sure there are no test leads connected to the Tester. Push the
TEST
button.
6.
After the
TEST
button is pressed, the voltage will ramp and hold. Set the
Voltage Adjust
control on the rear panel to the desired voltage.
7.
If tamperproof operation is desired, we suggest placing a calibration-style label over the
voltage adjust hole.
8.
Check the position of the
Ground Check
switch, and reset the test time using the
procedure
Adjustment of the High Voltage Test Time
.
Adjustment of the High Voltage Test Time
This procedure sets the length of time the HT-3000P will conduct the high voltage test.
The test time is specified by the safety agencies and is tied to the test voltage. Most safety
agencies will allow a much shorter test (usually 1 second vs. 1 minute) if the voltage is increased
by 20%. The HT-3000P is factory set for 2 seconds. Consult the safety agencies for the test time
for the type of equipment being tested. If a different test time is required, use this procedure to
set it.
1.
Defeat the
Ground Check
by turning the rear panel switch Off.
2.
Connect the HT-3000P to a correctly rated source of supply and turn the power switch to
the I or ON position. Push the
RESET
button. The yellow
TEST
indicator should light,
indicating that the HT-3000P is ready to test.
3.
On the rear panel turn the
Test Time
potentiometer. As soon as the potentiometer starts
turning, the meter will display “d” and the value can be set in 1 second increments from 1
to 60 seconds.
4.
If tamperproof settings are desired, the
Leakage Limit
control shaft may be pulled away
with a slight tug. (It may be reinserted for future adjustments.)
5.
Check the position of the
Ground Check
switch before conducting testing.
Setting the Ground Check Switch
This switch controls whether the Ground Test is conducted between the chassis and the
power supply cord grounding pin of the unit being tested. The safety agencies require that this
test be conducted on all equipment using a three-wire power supply cord. For other types of
equipment, such as some portable lamps and all double-insulated equipment, there are only two
wires in the power supply cord and the ground continuity check is waived by the safety agencies.
If the power supply cord of the equipment being tested has only two pins, the
Ground
Check
switch should be set to the Off position. When the
Ground Check
switch is turned Off,
both the green
Ground OK
and red
Ground Open
LED's on the front panel are lit as a reminder
to the operator that the ground continuity test is not being conducted.
If the power supply cord of the equipment being tested has three pins, the
Ground
Check
switch should be set to the On position.
When testing with the
Ground Check
switch Off, the operator must ensure that the
Return
lead is properly connected to exposed dead metal of the equipment being tested for
safety and to ensure that the HT-3000P properly reports all failures.
The small size of the
Ground Check
switch will allow a calibration-type adhesive label
to be placed over it to allow for tamper-proof operation.
Summary of Contents for HT-3000P
Page 2: ......
Page 4: ...2...
Page 14: ...11 Figure 2 Controls Indicators Connectors Model HT 3000P Front Panel...
Page 16: ...13 Figure 3 Controls Indicators Connectors Model HT 3000P Rear Panel...
Page 18: ...15...
Page 26: ...23...