background image

 

19 

3. 

Connect the HT-3000P to a correctly rated source of supply and turn the power switch to 
the I or ON position.  Push the 

RESET

 button.  The yellow 

TEST

 indicator should light, 

indicating that the HT-3000P is ready to test. 

4. 

Use the AC/DC button to select the type of output voltage.  

5. 

Make sure there are no test leads connected to the Tester.  Push the 

TEST

 button.  

6. 

After the 

TEST

 button is pressed, the voltage will ramp and hold.  Set the 

Voltage Adjust

 

control on the rear panel to the desired voltage. 

7. 

If  tamperproof  operation  is  desired,  we  suggest  placing  a  calibration-style  label  over  the 
voltage adjust hole. 

8. 

Check  the  position  of  the 

Ground  Check 

switch,  and  reset  the  test  time  using  the 

procedure 

Adjustment of the High Voltage Test Time

 

Adjustment of the High Voltage Test Time 

 

This  procedure  sets  the  length  of  time  the  HT-3000P  will  conduct  the  high  voltage  test.  

The  test  time  is  specified  by  the  safety  agencies  and  is  tied  to  the  test  voltage.    Most  safety 
agencies will allow a much shorter test (usually 1 second  vs. 1 minute) if the voltage is increased 
by 20%.  The HT-3000P is factory set for 2 seconds.  Consult the safety agencies for the test time 
for the type of equipment being tested.  If a different test time is required, use this procedure to 
set it.   
 
1. 

Defeat the 

Ground Check

 by turning the rear panel switch Off. 

2. 

Connect the HT-3000P to a correctly rated source of supply and turn the power switch to 
the I or ON position.  Push the 

RESET

 button.  The yellow 

TEST

 indicator should light, 

indicating that the HT-3000P is ready to test. 

3. 

On the rear panel turn the 

Test Time

 potentiometer.  As soon as the potentiometer starts 

turning, the meter will display “d” and the value can be set in 1 second increments from 1 
to 60 seconds. 

4. 

If  tamperproof  settings  are  desired,  the 

Leakage  Limit

  control  shaft  may  be  pulled away 

with a slight tug.  (It may be reinserted for future adjustments.) 

5. 

Check the position of the 

Ground Check

 switch before conducting testing. 

 

Setting the Ground Check Switch 

 
 

This switch controls whether the Ground Test is conducted between the chassis and the 

power supply cord grounding pin of the unit being tested.  The safety agencies require that this 
test  be  conducted  on  all  equipment  using  a  three-wire  power  supply  cord.    For  other  types  of 
equipment, such as some portable lamps and all double-insulated equipment, there are only two 
wires in the power supply cord and the ground continuity check is waived by the safety agencies. 
 

If the power supply cord of the equipment being tested has only two pins, the 

Ground 

Check

 switch should be set to the Off position.  When the 

Ground Check

 switch is turned Off, 

both the green 

Ground OK

 and red 

Ground Open

 LED's on the front panel are lit as a reminder 

to the operator that the ground continuity test is not being conducted. 
 

If  the  power  supply  cord  of  the  equipment  being  tested  has  three  pins,  the 

Ground 

Check

 switch should be set to the On position. 

 

When  testing  with  the 

Ground  Check

  switch  Off,  the  operator  must  ensure  that  the 

Return

  lead  is  properly  connected  to  exposed  dead  metal  of  the  equipment  being  tested  for 

safety and to ensure that the HT-3000P properly reports all failures. 
 

The small size of the 

Ground Check

 switch will allow a calibration-type adhesive label 

to be placed over it to allow for tamper-proof operation. 

Summary of Contents for HT-3000P

Page 1: ...HT 3000P Dielectric Withstand Tester 0 2000 Volts AC Output 0 2800 Volts DC Output Instruction Manual COMPLIANCE WESTUSA...

Page 2: ......

Page 3: ...logic controlled circuit design and ergonomic front panel and represents the latest in high voltage production line testing To fully appreciate all the features of your new meter we suggest that you t...

Page 4: ...2...

Page 5: ...and Rear Panel Features 11 Initial Checkout Procedure 16 Setting up the HT 3000P for Production Line Testing 17 Factory Settings 17 Display of Leakage Limit and Duration settings 17 Adjustment of the...

Page 6: ...ndamentals clearly and be aware that high voltage is adept and creative at completing a path to ground Instructions should include a warning against any metal jewelry Operators should not allow others...

Page 7: ...tion between the operator and the test being conducted The HT 3000P is designed to bleed the high voltage away after the test has concluded In order to ensure that any voltage present in the equipment...

Page 8: ...continuously during the test When testing with the Ground Check switch in the Off position the operator must ensure that the Test Return Lead is properly connected to exposed dead metal of the equipme...

Page 9: ...s is due to the AC voltage charging all capacitors connected between primary voltage and ground If the overall value of these capacitors cause a leakage current of more than 20 mA to flow at the desir...

Page 10: ...ntil the voltage has discharged to a safe level to protect the operator and also to ensure that there is no energy stored in the tested equipment Testing Equipment with non Standard Plugs or Pigtail O...

Page 11: ...ff by a resistor bank in the HT 3000P The voltage discharge progress is shown by the front panel meter Failure modes are shown by the front panel LED s for quick troubleshooting Convenience and testin...

Page 12: ...available by removing adjustment shafts ENVIRONMENTAL Operating Temperature 15 40 C Relative Humidity Range 0 90 non condensing GENERAL Input power requirements See rear panel for input voltage and f...

Page 13: ...ock absorbing material on all sides of the container Remove the tester from its container and place it on a test bench AC Line Voltage Requirements AC line voltage requirements for your Tester are not...

Page 14: ...11 Figure 2 Controls Indicators Connectors Model HT 3000P Front Panel...

Page 15: ...d the test will be terminated 5 Full Voltage LED If full voltage is successfully reached the Full Voltage LED will light and the high voltage duration timer starts 6 Hipot Fail LED Indicates failure o...

Page 16: ...13 Figure 3 Controls Indicators Connectors Model HT 3000P Rear Panel...

Page 17: ...ent of the High Voltage Ramp Time for this procedure 4 Test Time Adjustment Used to adjust high voltage test duration See Adjustment of the High Voltage Test Duration section for this procedure 5 Leak...

Page 18: ...15...

Page 19: ...osition 8 Verify that all test leads are disconnected from the HT 3000P 9 Push the red RESET button then the yellow TEST button 10 The tester should sound the buzzer The red Ground Open and red RESET...

Page 20: ...may light and in some instances the Ground OK LED may extinguish Passage of these tests indicates that the HT 3000P is functioning properly and that it is safe to use If the results of the performanc...

Page 21: ...ons Use this procedure if adjustment to a different ramp time within the range of 1 5 seconds is required 1 Defeat the Ground Check by turning the rear panel switch to the Off position 2 Connect the H...

Page 22: ...the rear panel turn the Test Time potentiometer As soon as the potentiometer starts turning the meter will display d and the value can be set in 1 second increments from 1 to 60 seconds 4 If tamperpro...

Page 23: ...e power switch to the I or ON position Push the RESET button The yellow TEST indicator should light indicating that the HT 3000P is ready to test 2 Connect the HTT 1 to the HT 3000P tester using the c...

Page 24: ...f time set in the Test Duration procedure If a breakdown of the insulation system between primary voltage and ground is detected the red Hipot Fail LED will light the buzzer will sound and the test wi...

Page 25: ...ssistance Phone 800 748 6224 Technical Assistance is available from Compliance West USA between the hours of 8 30 AM and 4 30 PM Pacific Time Compliance West USA 2120 Jimmy Durante Blvd Suite 118 Del...

Page 26: ...23...

Page 27: ...ad with a banana plug on one end are required to conduct the performance test The test equipment required for the calibration procedure is a DMM able to read true rms 0 2000 Vac 1 0 2800 Vdc 1 and a 3...

Page 28: ...the front panel and case with a mild solution of detergent and a damp sponge Clean dust from the PWB with clean dry low pressure 20 psi Calibration Procedure The Calibration Procedure should be perfo...

Page 29: ...r clip end to the 3 ohm resistor 3 Connect the banana plug end of the other lead to the round grounding jack of the DEVICE receptacle and the alligator clip end to the 3 ohm resistor 4 Turn the Ground...

Page 30: ...a test and using the 1000 1 voltage probe to check the meter at different voltages If the voltages are not within 2 at each voltage the voltage calibration must be performed again starting at step 1 a...

Reviews: