10
ELECTRICAL
Output
0-2000 Volts AC or 0-2800 Volts DC .
Pigtail and Receptacle adapters available for most
configurations. Consult factory.
Leakage Current
0-20 mA AC; 0-5 mA DC
Pass/Fail Criteria:
Continuity:
3 ohms ± 5% * (defeatable)
Leakage Current:
Pass/Fail point user adjustable.
Dielectric Breakdown:
Separate high frequency detection circuit for
breakdown spike detection
Test Time:
User adjustable 1->60 sec.
Voltage Ramp-up Time:
User adjustable 1-5 sec.
Voltage Ramp-down Time:
Factory set 1 sec. approx.
Pass/Fail Repeatability
±
3%
Meter accuracy
+ 2% from 500-2000 volts AC
+ 2% from 500-2800 volts DC
Duty cycle
100 %
Test adjustments
Rear Panel:
Ramp Time
Test Time
Leakage Limit
Voltage Adjust
Tamperproof operation available by removing
adjustment shafts.
ENVIRONMENTAL
Operating Temperature
15-40
°
C
Relative Humidity Range
0-90% non-condensing
GENERAL
Input power requirements
See rear panel for input voltage and fuse ratings
Weight
10½ lbs.
SAFETY AGENCY TOPICS
Transformer Output
< 500VA
Visual Indication of Voltage Output
Provided by front panel meter, directly connected to
high voltage output
Failure Indication
Audible, provided by internal buzzer
Visual, provided by red LEDs on front panel
Test automatically terminated on failure
Continuity Test
Provided; 3 ohm pass/fail point*. May be defeated
for testing of products with two wire power supply
cords.
Leakage Test
Provided; 5 mA AC/5 mA DC factory set pass/fail
point, user adjustable.
Operation Instructions
Provided on rear panel.
Table 1. HT-3000P Specifications
*Ground Continuity Pass/Fail Level can be adjusted from .5 to 5 ohms during the calibration procedure.