Annex 1: Selected error messages of the device
Error
Description and debugging
Error 1
A/D converter is at low limit, 0x0000.
Contact support.
Error 2
Temperature probe is not connected or cable is damaged.
Error 3
Internal converter communication error.
Contact support.
Error 5
Few valid measured values for calculating the average value
for the recording interval (more than 1/8 of invalid values were
measured from the given recording interval).
Error 9
Measured value is out of range or probe is shorted.
Error 10
The device is switched off, measured values are not available.
Error 11
No valid data of the counter, batteries missing in the device.
Reset the counter content by means of the SW, repeat
the measurement.
Error 13
Error 14
Error 15
CO
2
concentration sensor is not available, or it does not work correctly.
Have the device repaired. This applies to the CO
2
concentration
sensor devices only.
Error 16
The measured CO
2
value is not available due to low battery voltage.
Recharge the battery.
Error 17
The digital T / RH probe does not communicate with the instrument.
Check its connection.
Error 18
The digital T/RH probe memory does not communicate with the
instrument.
Contact support
.
Error 19
Measured value is not available.
In device U3121 check, and, if needed, replace the Digi/E probe
connected. In the other models a device failure may be suspected.
Summary of Contents for U0110
Page 5: ...IE LGR Uxxxx 16 5 General view of a Uxxxx series datalogger U0141 model...
Page 64: ...64 IE LGR Uxxxx 16 Dimensions U0110 U0111 U0122...
Page 65: ...IE LGR Uxxxx 16 65 U0121 U0141...
Page 66: ...66 IE LGR Uxxxx 16 U0246 U0141T U0541 U5841 U6841 U7844 U2422...
Page 67: ...IE LGR Uxxxx 16 67 U3121 U3120 U4130...
Page 68: ...68 IE LGR Uxxxx 16 U3430 U4440 U3631...
Page 69: ...IE LGR Uxxxx 16 69 U8410...