IE-LGR-Uxxxx-16
33
Channels
_______________________________________
In this panel you can set all input channels. The assignment
of the measurement value and its range to a channel is factory-set
and cannot be changed.
Depending on the device type, up to three calculated channels are
available (for details see
).
Supply a suitable name of the location to be measured (or calculated
value) for each channel and decide whether it will be switched on for
measurement and for recording.
Further items, like those for entering the decimal place number,
the name of the physical unit and the conversion of the measurement
values by means of the so-called
User calibration
are available
in channels with voltage and current inputs only. In these inputs
no physical quantity unit is
displayed on the device’s LCD display.
If the value cannot be displayed on the device's display for a large
number of digits, then Hi (if greater than the displayable number) or Lo
(if less than the displayable number) is displayed. However, this error
message only concerns the display on the device display,
not the measurement and recording.
An example of user-defined calibration for a sensor having a 4 to
20
mA current output. The sensor is connected to the datalogger’s
current input; the sensor’s measurement range is -30 to +80 °C:
Lower point A: Input value: 4 will be shown as -30
Upper point B: Input value: 20 will be shown as 80
In a channel for external Pt 1000 probes there exists the possibility to
enter the probe parameters to correct the measurement error caused
by probe cable length.
In binary channels you can decide whether a voltage signal or a contact
(an open-collector transistor) will be connected to the input. Then enter
textual descriptions of the states for both input levels. In this way
the states will be displayed in the record and in the Online display.
At the same time, from the menu, choose symbols for displaying these
states by the device.
For a counter channel, decide whether to report an error or continue
from zero after it is overflowed. . The range is 24 bits, it means max.
16 777 215.
In the next step, set alarm states for each measured variable. You can
set up to 2 alarms for each measured quantity (alarm 1 and alarm
2 buttons). Each alarm must be enabled (on/off) first.
Furthermore, choose whether the alarm should arise from exceeding
the limit value or from dropping under it (
value is greater than / value is
Summary of Contents for U0110
Page 5: ...IE LGR Uxxxx 16 5 General view of a Uxxxx series datalogger U0141 model...
Page 64: ...64 IE LGR Uxxxx 16 Dimensions U0110 U0111 U0122...
Page 65: ...IE LGR Uxxxx 16 65 U0121 U0141...
Page 66: ...66 IE LGR Uxxxx 16 U0246 U0141T U0541 U5841 U6841 U7844 U2422...
Page 67: ...IE LGR Uxxxx 16 67 U3121 U3120 U4130...
Page 68: ...68 IE LGR Uxxxx 16 U3430 U4440 U3631...
Page 69: ...IE LGR Uxxxx 16 69 U8410...