Response time (air flow 1 m/s approx.)
(*2)
:
Temperature: t
63
< 2 min, t
90
< 8 min (temperature step
20 °C)
Relative humidity: t
63
< 45 s, t
90
< 4 min, t
99
<4 h.
(Measured at constant temperature for 30%
humidity step)
Barometric pressure: t
90
<44s
Display resolution:
Temperature incl. dew-
point temperature: 0.1 °C
Relative humidity: 0.1 %RH
Barometric pressure: 1 hPa
Recommended calibration interval:
1 year
(*2)
Important information is provided in the chapter
“Models produced – U4130”
U4440
_________________________________________
Measured values:
Internal temperature, relative humidity, barometric
pressure and the concentration of CO
2
in the air.
Dew-point temperature calculated from the internal
temperature and relative humidity.
Ranges:
Temperature: (-
20 to +60) °C
Relative humidity: (0 to 100) %RH without permanent
condensation
(*2)
.
Barometric pressure (absolute): 700 hPa to 1100 hPa
Concentration of CO
2
in the air: (0 to 5000) ppm (range
(0 to 10000) ppm on request)
Dew-point temperature: (-60 to +60)
°C
Accuracy
(*1)
:
Temperature: ± 0.4 °C
Relative humidity:
-
sensor accuracy ±1.8 %RH at 23 ºC in the
range of (0 to 90) %RH
-
hysteresis < ±1 %RH
- non-
linearity < ±1 %RH
- temperature error: see diagrams in
Barometric pressure: ± 1.3 hPa at 23 °C
CO
2
concentration in the air:
50 + 0.03 × MV [ppm CO
2
at 23 °C and 1013 hPa]
Temperature dependency in range (-20...45)
°C:
typ. ± (1 + MV / 1000) [ppm CO
2
/°C]
MV…measured value
Summary of Contents for U0110
Page 5: ...IE LGR Uxxxx 16 5 General view of a Uxxxx series datalogger U0141 model...
Page 64: ...64 IE LGR Uxxxx 16 Dimensions U0110 U0111 U0122...
Page 65: ...IE LGR Uxxxx 16 65 U0121 U0141...
Page 66: ...66 IE LGR Uxxxx 16 U0246 U0141T U0541 U5841 U6841 U7844 U2422...
Page 67: ...IE LGR Uxxxx 16 67 U3121 U3120 U4130...
Page 68: ...68 IE LGR Uxxxx 16 U3430 U4440 U3631...
Page 69: ...IE LGR Uxxxx 16 69 U8410...