58
IE-LGR-Uxxxx-16
Dew-
point temperature: ±1.5 °C at an ambient
temperature of T< 25
°C and RH >30 %.
For details see diagrams in
(*1)
During battery charging temporary measurement
inaccuracy can occur.
Response time (air flow 1 m/s approx.)
(*2)
:
Temperature: t
63
< 2 min, t
90
< 8 min (temperature step
20 °C)
Relative humidity: t
63
< 45 s, t
90
< 4 min, t
99
<4 h.
(Measured at constant temperature for 30%
humidity step)
Barometric pressure: t
90
<44s
CO
2
concentration in the air: according to the set CO
2
Measuring interval (2 min / 10 min)
(*2)
Important information is provided in the chapter
“Models produced – U4440”
Display resolution:
Temperature incl. dew-
point temperature: 0.1 °C
Relative humidity: 0.1 %RH
Barometric pressure: 1 hPa
CO
2
concentration in air: 1 ppm
Recommended calibration interval:
1 year
U5841
_________________________________________
Measured values:
3 x voltage input (0 - 10) V
1 x binary input. This input can be configured to voltage
or a dry contact (with open collector transistor
acceptance).
This
input
can
capture
connected/disconnected voltage status.
Range and input levels:
Voltage inputs:
Range: (0 to 10) V DC
Accuracy: ± 10 mV
Input resistance: approx. 130 k
Ω
Binary input configured for voltage signal:
Input voltage for the „L“ level: < 0.8 V
(*)
Input voltage for the „H“ level: > 2 V
Minimum applicable voltage: 0 V
Maximum applicable voltage: +30 V DC
(*)
When the input is disconnected, the device will
measure the „L“ level.
Summary of Contents for U0110
Page 5: ...IE LGR Uxxxx 16 5 General view of a Uxxxx series datalogger U0141 model...
Page 64: ...64 IE LGR Uxxxx 16 Dimensions U0110 U0111 U0122...
Page 65: ...IE LGR Uxxxx 16 65 U0121 U0141...
Page 66: ...66 IE LGR Uxxxx 16 U0246 U0141T U0541 U5841 U6841 U7844 U2422...
Page 67: ...IE LGR Uxxxx 16 67 U3121 U3120 U4130...
Page 68: ...68 IE LGR Uxxxx 16 U3430 U4440 U3631...
Page 69: ...IE LGR Uxxxx 16 69 U8410...