26
IE-LGR-Uxxxx-16
U6841
Datalogger with 3 inputs 0 - 20 mA
and 1 binary input
This model is used to measure up to three signals
from current loops (0 to 20) mA DC while allowing one
binary input to be monitored. The current inputs are
passive and are not galvanically isolated from one
another. This means that the connected sensors
cannot
be powered by the current loop and the negative
terminals ("-") are connected to each other. These facts
must be taken into consideration when designing
the device into the circuit. Incorrect connection may
cause degradation of the measured values. Standard
outputs (4 to 20) mA can also be measured with current
inputs without any problems.
Binary input serves to monitor voltage signals or signals
coming from a dry contact. The binary input changes are
registered immediately, irrespectively of the recording
interval setting. To safe record the input status, the signal
must be present for at least 1 s. This input can capture
connected/disconnected voltage status.
Input signals are connected to a pluggable terminal
block. The
device’s ingress protection is IP 20. Due to its
low ingress protection the device is not suitable for using
in dusty rooms or rooms exposed to water ingress.
The maximum length of the input wires should not exceed
30 m. It is recommended to use shielded cables.
The battery in the device cannot be charged.
Connection:
The procedure for connecting the wires to the terminals
can be found in
Summary of Contents for U0110
Page 5: ...IE LGR Uxxxx 16 5 General view of a Uxxxx series datalogger U0141 model...
Page 64: ...64 IE LGR Uxxxx 16 Dimensions U0110 U0111 U0122...
Page 65: ...IE LGR Uxxxx 16 65 U0121 U0141...
Page 66: ...66 IE LGR Uxxxx 16 U0246 U0141T U0541 U5841 U6841 U7844 U2422...
Page 67: ...IE LGR Uxxxx 16 67 U3121 U3120 U4130...
Page 68: ...68 IE LGR Uxxxx 16 U3430 U4440 U3631...
Page 69: ...IE LGR Uxxxx 16 69 U8410...