Settings / System
Settings / System
62
ColorLite GmbH | sph860/sph900
On the top level menu
“setting”
and
“memory”
are no longer se-
lectable. Furthermore, it is not pos-
sible to measure a “Standard”.
Standards can only be loaded from
the memory.
To disable the
“User mode”
select
“Measure”
from the main menu.
In the following window go on
“Exit user mode”
and confirm this
with Enter. You will be prompted to
enter the password.
After entering the correct password, the
“User Mode”
is
canceled and the unit will return to the measurement
menu.
If the
“User mode”
and also the function
“Use
Working-Standard”
[6.1.7
Using a work-
ing standard]
are activated, it is only possible to
calibrate with the
Working-Standard
.
6.1.6 Power supply
An optional device holder is available for the
sph860/sph900 which has an integrated power supply or a
separate power supply can be ordered.
When using an external power supply the Auto-Power–Off
function is deactivated
[see 4.6
Auto-Power-Off]
.
In the menu
“Settings”
-
“System”
go to
"Ext. power
supply"
. By pressing the
“Enter”
_
key the mode is
activated, display by a tick in the check box. By pressing
the same key the function is deactivated
If the rechargeable battery is used, switch off the
ext. power supply mode.