Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
1-8
System bus (CAN)
Standard
Input Specification
AC input voltage
3phase , 3 wire + ground
17
200/220 Vac (operating range 180 -242 Vac)
380/400 Vac (operating range 342 - 440 Vac)
440/480 Vac (operating range 396 - 528 Vac)
*
*
Call for Availability
AC frequency range
47-63 Hz
Power factor
0.6 (200/220Vac)
0.6 (380/400Vac)
0.6 (440/480Vac)
General Specification
Maximum Remote Sense
Line Drop Compensation
2% of full scale voltage per line (4% total)
Weight
< 35 kg / 77.16 lbs.
Dimensions (HxWxD) mm
18
132.8 x 428 x 610 mm / 5.23 x 16.85 x 24.02 inch
Operating Temperature
Rage
0°C ~ 40°C
Storage Temperature Rage
-40°C ~ +85°C
Approval
CE
All specifications are subject to change without prior notice.
Note
1.
Minimum output voltage <0.5% of rate voltage. The 62020H-150S minimum
output rated voltage is 1.5V.
2.
Minimum output current <0.2% of rate current.
3.
The Range applicable for change only valid in CV/CC MODE and the Range
of TABLE MODE and SAS MODE is Full Scale.
4.
The measurement frequency range is 20k Hz ~ 20M Hz.
5.
The output voltage range is from 10% to 100% and the output current is
measured under the condition of full load.
6.
This setting is only valid when there is output and the voltage as well as the
current setting is larger than the one specified in
Note 1
and the load current
is 40% over Imax. When the output is connected to capacitor, the voltage
slew rate will decrease if the capacitance is increased.
7.
This setting is valid only when the load voltage is larger than the one
specified in
Note 1
. The factory default is INF. Be sure to adjust the slew
rate settings as required.
8.
10% variation under rated voltage.
9.
For 0-100% load step with nominal line voltage.
10. Under the maximum output power condition of rated voltage.
11. The maximum drift of output power during 30 minutes test period when the
input, loading and ambient temperature are fixed.
12. The maximum drift of output power after warmed up for 30 minutes and 8
hours test period when the input, loading and ambient temperature are fixed.
13. The change caused by the ambient temperature per centigrade when the
input and loading are fixed.
14. Over 50% of maximum output voltage and the
loading slew rate is 1A/us for
rise and fall.
15. Either Ethernet or GPIB can be selected when shipping.
16. Please consult with the manufacturer when there are 5 DC Power Supplies
are connected in parallel. There is parallel mode for DC Power Supply when
the I-V Curve function is enabled.
17. Varied by local voltage regulation, all models of 5KW, 10kW & 15kW in the
62000H Series with Solar Array Simulation have 200/220 Vac, 380/400 Vac