Section 2: Installation and Setup
D4K35 Setup Guide
2-7
020-100855-02 Rev. 1 (09-2014)
c. Insert the LPS support into the rack stand and align the pins with the holes.
d. Lift the LPS into the frame.
e. Remove the bottom 10 panels from the back of the rack stand.
f. Assemble the three piece airflow duct (supplied with LPS). See
Figure 2-4
. First snap the 2 “L”
shaped parts together, and then snap this into the 6 stand out pins of the third piece (flat panel)..
g. Position the airflow duct directly behind the LPS.
CAUTION!
Do not install any components in front
of the lamp power supply that will restrict airflow..
See
Figure 2-5
.
h. Insert airflow duct into rack stand from the outside.
See
Figure 2-4.
i. Secure airflow duct to rack stand using screws from
the rack stand panels.
j. When installing other components in the rack stand
allow at least 2” clearance above the LPS handles for
ease of servicing. See
Figure 2-5.
k. Wire all components within the rack stand and
replace the side panel.
l. Twist LPS cable 12 twist/meter.
Figure 2-4 Steps to Installing LPS
Insert support frame
Lift LPS into frame
2.
Snap 3 piece airflow duct
together using stand out pins
3.
1.
2”
Keep
Clear
Figure 2-5Steps to Installing LPS
Summary of Contents for D4K35
Page 1: ...D4K35 S e t u p G u i d e 020 100855 02...
Page 2: ......
Page 3: ...D4K35 S e t u p G u i d e 020 100855 02...
Page 8: ......
Page 12: ......
Page 40: ......
Page 48: ......
Page 50: ......
Page 64: ......
Page 65: ......