Rotational Vacuum Concentrator RVC 2-33 CDplus
with infrared heating
8 Maintenance and service
Version 08/2009, Rev. 3.6 of 22/11/2021
• sb-bs
79 / 105
Translation of the original operating manual
Fig. 69: Chamber and rotor holder to be cleaned
Installing the rotor holder and rotor-side magnetic coupling
WARNING
Risk of crushing when installing the magnetic coupling
The magnets of the magnetic coupling and the ones of the motor attract
one another (red, see the illustration) so that there is a risk of crushing
your fingers between the components!
• When installing the magnetic coupling, do not put your fingers under
the magnetic coupling!
Fig. 70: Installing the magnetic coupling
• Grab the magnetic coupling by its sides and install it in the rotor
chamber.
• Install the parts and assemblies in reverse order with regard to their
removal.
• Connect the mains power connector and switch the unit on via the
mains power switch.
• Perform a test run.