Rotational Vacuum Concentrator RVC 2-33 CDplus
with infrared heating
8 Maintenance and service
78 / 105
Version 08/2009, Rev. 3.6 of 22/11/2021
• sb-bs
Translation of the original operating manual
• Loosen all four fastening screws.
Fig. 67: Loosening the fastening screws
• Remove the magnetic coupling by lifting it up and off the rotor holder.
Fig. 68: Removing the magnetic coupling
• Then, the rotor chamber, rotor holder and magnetic coupling can be
cleaned.
CAUTION
Ensure that cleaning agents or disinfectants cannot flow into the opening
at the bottom of the rotor chamber (see the following illustration)!
This may lead to irreversible damage in the motor chamber.
• Wipe the rotor chamber, rotor holder and magnetic coupling clean or
clean them with water and soap or a disinfectant.