7 Malfunctions and error correction
72 / 105
Version 08/2009, Rev. 3.6 of 22/11/2021
• sb-bs
Translation of the original operating manual
Pos : 181 /200 Chris t/370 R VC-BA (ST ANDARDM ODU LE)/ 070 St örungen und Fehlers uc he/070-0010-0050 Klei nflansc hv erbi ndungen @ 25\m od_1404985190199_68.docx @ 187485 @ 3 @ 1
→
7.1.5
Small flange connections
Leakages are often due to improper small flange connections between the
various components and hose connections or to leakages in the valves.
• Loosen the connection and place the centring ring (with sealing ring
inside) in a centred manner between the flange connections.
• Seal the connection with the clamping ring or clamping chain by
tightening the wing nut.
• Ensure that the centring ring neither slips out of place nor gets jammed.
Fig. 59: Small flange and centring ring
Fig. 60: Small flange with centring ring
and small flange
Fig. 61: Attaching the clamping ring
Fig. 62: Tightened clamping ring