Rotational Vacuum Concentrator RVC 2-33 CDplus
with infrared heating
6 Operation
Version 08/2009, Rev. 3.6 of 22/11/2021
• sb-bs
45 / 105
Translation of the original operating manual
Evaporation with low capacity
The tubes must be installed symmetrically so that the buckets and their
inserts are loaded evenly to avoid imbalance.
correct
incorrect
Fig. 15: Permissible loading of swing-out rotors with low capacity
Pos : 140 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 141 /200 Chris t/370 R VC-BA (ST ANDARDM ODU LE)/ 060 Betrieb/ 060-0040- 0030-0010 G ef äß e (auß er 2-18) @ 25\m od_1404985119757_68. docx @ 187205 @ 4 @ 1
→
6.4.3.1
Tubes
• Load the tubes outside of the rotational vacuum concentrator. Solvents
in the buckets or rotor blocks cause corrosion.
• Fill the tubes carefully and arrange them according to their weight.
• Follow the safety instructions and hazard warnings (see chapter 3 -