Rotational Vacuum Concentrator RVC 2-33 CDplus
with infrared heating
5 Set-up and connection
Version 08/2009, Rev. 3.6 of 22/11/2021
• sb-bs
37 / 105
Translation of the original operating manual
3.
Connection for aeration and air injection with inert gas
1
Aeration hose connection
2
Air injection hose nozzle
3
Fastener for non-required connection element; here: air filter
CAUTION
If gas (e.g. inert gas) is supplied through the hose nozzle for aeration or
for “air injection”, the threads of the hose nozzles must be sealed with
thread sealant, e.g. Loctite® 542 (Loctite® part number: 54223, bottle with
10ml).
Pos : 113 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 114 /200 Chris t/370 R VC-BA (ST ANDARDM ODU LE)/ 050 Aufst ellung und Anschl uss /050- 0040 Vak uumv erbi ndungen (auß er 2-18 HC L) @ 25\m od_1404985087233_68. docx @ 187051 @ 2 @ 1
→
5.4
Vacuum connections
The vacuum connection is realised by way of a vacuum hose with standard
flange connections, clamping rings, or chains, and centring rings.
NOTE
The small flange connections must be installed correctly in order to
prevent leaks (see chapter 7.1.5 - "Small flange connections ").
1
Centring ring
2
Clamping ring
3
Flange connection
Fig. 8: Connecting pieces for vacuum connection