Rotational Vacuum Concentrator RVC 2-33 CDplus
with infrared heating
3 Safety
28 / 105
Version 08/2009, Rev. 3.6 of 22/11/2021
• sb-bs
Translation of the original operating manual
Pos : 73 / 200 C hrist/ 370 R VC-BA ( STAND ARDMODU LE) /030 Sic herheit/ 030-0080 Sic her heit sei nricht ungen------------------------------------------------------ @ 86\m od_1635952536376_68. docx @ 906388 @ 2 @ 1
→
3.8
Safety devices
WARNING
The safety devices have been installed to protect the personnel using the
unit against injury.
If all of the safety devices are not fully functional
and
if all of the operating
elements are not freely accessible, serious health damage may result.
• The safety-relevant operating elements (mains power switch) must be
accessible at all times!
• Do not manipulate, remove or disable the safety devices or guards in
any way!
• The correct operation of the safety devices must be checked at regular
intervals in accordance with the applicable national and international
laws, rules and regulations concerning health and safety and the
prevention of accidents!
Pos : 74 / 010 U nivers almodule/ Leerz eile @ 0\m od_1202116244500_0. doc x @ 114 @ @ 1
Pos : 75 / 200 C hrist/ 370 R VC-BA ( STAND ARDMODU LE) /030 Sic herheit/ 030-0080- 0020 Sy stem kontr olle @ 25\mod_1404985015206_68. doc x @ 186883 @ 3 @ 1
→
3.8.1
System check
An internal system check monitors the data transfer and sensor signals with
regard to plausibility. Errors are detected by continuous self-monitoring of
the system. Error messages are displayed in the process & equipment info
menu (see (see chapter 6.5.3.4 - "Process and equipment information") or
(see chapter 7.1.1 - "Error messages")).
Pos : 76 / 010 U nivers almodule/ Leerz eile @ 0\m od_1202116244500_0. doc x @ 114 @ @ 1
Pos : 77 / 200 C hrist/ 370 R VC-BA ( STAND ARDMODU LE) /030 Sic herheit/ 030-0080- 0030 Netzaus fallsic her hei t (auß er 2- 18) @ 25\mod_1404985016253_68. docx @ 186897 @ 3 @ 1
→
3.8.2
Power failure safety
The unit stores the current process parameters cyclically. In the event of a
power failure, the electromagnetic stop valve will close automatically, the
rotor chamber will be aerated by the aeration valve and the rotor
decelerates brakeless. After the restart of the power supply, the
evaporation process continues with the last stored parameters.
Pos : 78 / 010 U nivers almodule/ Leerz eile @ 0\m od_1202116244500_0. doc x @ 114 @ @ 1
Pos : 79 / 200 C hrist/ 370 R VC-BA ( STAND ARDMODU LE) /030 Sic herheit/ 030-0080- 0040 Dec kelv erriegel ung @ 25\mod_1404985013844_68. doc x @ 186869 @ 3 @ 1
→
3.8.3
Lid lock device
The RVC can only be started when the lid is properly closed. The lid can
only be opened when the rotor has stopped. If the lid is open, the RVC
cannot be started.
Pos : 80 / 010 U nivers almodule/ Leerz eile @ 0\m od_1202116244500_0. doc x @ 114 @ @ 1
Pos : 81 / 200 C hrist/ 370 R VC-BA ( STAND ARDMODU LE) /030 Sic herheit/ 030-0080- 0040 Sc hutzl eiter prüfung @ 25\m od_1404985017329_68. docx @ 186911 @ 3 @ 1
→
3.8.4
Earth conductor check
For the earth conductor check, there is an equipotential bonding screw on
the rear panel of the RVC. An earth conductor check can be carried out
with the aid of a suitable measuring instrument.