Rotational Vacuum Concentrator RVC 2-33 CDplus
with infrared heating
7 Malfunctions and error correction
Version 08/2009, Rev. 3.6 of 22/11/2021
• sb-bs
63 / 105
Translation of the original operating manual
Pos : 167 /200 Chris t/370 R VC-BA (ST ANDARDM ODU LE)/ 070 St örungen und Fehlers uc he/070 St örungen und Fehlers uc he=========================== @ 25\m od_1404985185894_68.docx @ 187415 @ 1 @ 1
→
7
Malfunctions and error correction
Pos : 168 /200 Chris t/370 R VC-BA (ST ANDARDM ODU LE)/ 070 St örungen und Fehlers uc he/070-0010 Allgem eine St örungen @ 25\m od_1404985186669_68.docx @ 187429 @ 2 @ 1
→
7.1
General malfunctions
Malfunctions are displayed in the menu "Process & equipment info" (see
chapter 6.5.3.4 - "Process and equipment information"). A sound signal
sounds when an error message is generated.
• Eliminate the source of the problem (see the following chapters).
• Acknowledge the error messages by pressing the right-hand function
key.
Pos : 169 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 170 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 171 /200 Chris t/370 R VC-BA (ST ANDARDM ODU LE)/ 070 St örungen und Fehlers uc he/070-0010-0010 F ehl ermel dungen ( auß er 2-18) @ 25\m od_1404985187671_68.docx @ 187443 @ 3 @ 1
→
7.1.1
Error messages
Error
Possible reason
Correction
No vacuum sensor signal
•
The control unit receives an
invalid value from the vacuum
sensor.
•
Check the connecting cables
and plug-in connections of the
vacuum sensor. If necessary,
replace the vacuum sensor.
No rotor chamber sensor signal
•
The temperature sensor of the
rotor chamber is disconnected
or defective
•
Contact the after-sales service
(see chapter 7.2 - "Service
contact")
No rotor sensor signal
•
The rotor temperature sensor is
disconnected or defective
•
Contact the after-sales service
(see chapter 7.2 - "Service
contact")
No ice condenser sensor signal
•
The data communication link to
the cooling trap is interrupted,
or the ice condenser
temperature sensor is
disconnected or defective
•
Check the data link between the
rotational vacuum concentrator
and the cooling trap
Lid cannot be opened
•
The rotor chamber is not
aerated completely
•
Aerate the rotor chamber
completely
Lid contacts error
•
The system has detected an
open lid during evaporation.
The rotor will be stopped
immediately. The lid must be
closed while the rotor is
moving.
•
Check the lid contacts. If
necessary, contact the after-
sales service (see chapter 7.2 -
"Service contact")
Rotor overtemperature
•
The maximum permissible rotor
temperature of 85°C has been
exceeded
•
If the system does not cool
down, disconnect it immediately
from the power supply. Contact
the after-sales service (see
chapter 7.2 - "Service contact")
Chamber overtemperature
•
The maximum permissible
chamber temperature has been
exceeded
•
If the system does not cool
down, disconnect it immediately
from the power supply. Contact
the after-sales service (see
chapter 7.2 - "Service contact")