Rotational Vacuum Concentrator RVC 2-33 CDplus
with infrared heating
3 Safety
Version 08/2009, Rev. 3.6 of 22/11/2021
• sb-bs
27 / 105
Translation of the original operating manual
3.7.9
Hazards due to the evaporation of harmful precursors
If infectious, toxic, pathogenic or radioactive substances are to be
evaporated, the user is responsible for ensuring that all applicable safety
regulations, guidelines, precautions and safety measures are observed.
DANGER
Risk of poisoning/infection from the starting products!
• Infectious, toxic, pathogenic and radioactive substances may only be
evaporated in suitable vessels. For your own protection, it is essential
to take appropriate precautions!
• Do not evaporate corrosive raw materials (especially acidic
substances) without special protective measures and equipment (e.g.
additional cold trap to protect the vacuum pump). It is essential to
consult the manufacturer (see chapter 7.2 - "Service contact").
DANGER
Explosion hazard due to azides!
• Special care must be taken when handling azides, as a dangerous
explosive is formed in combination with copper or non-ferrous metals!
It is essential to consult the manufacturer (see chapter 7.2 - "Service
contact").
• Local measures for the containment of harmful emissions must be
observed (depending on the substances to be dried).
• Heat-resistant gloves are required as protective clothing for operating
the rotary vacuum concentrator. The material to be dried may require
further safety measures (e.g. drying of infectious, toxic, radioactive or
pathogenic substances).).