Visual 

inspection 

   

Check the condition and function of the product regularly. Check to ensure that material is free from damage. If there are 
signs of wear, the product must be inspected and serviced.

Always read the manuals for all assistive devices used during a transfer.
Keep the manual where it is accessible to users of the product.
Do not leave the patient unattended during a transfer situation.
The product should be inspected yearly, and otherwise as required, by a qualified technician. Service must be performed 
by a qualified technician.

Functional inspection

Always read the manual

Art. no.

7600

Manual no: 720-En  Ver. 1  150617

SWL: 205kg/450 lbs

ReTurn7600

System

RoMedic

TM

Summary of Contents for ReTurn7600

Page 1: ... all assistive devices used during a transfer Keep the manual where it is accessible to users of the product Do not leave the patient unattended during a transfer situation The product should be inspected yearly and otherwise as required by a qualified technician Service must be performed by a qualified technician Functional inspection Always read the manual Art no 7600 Manual no 720 En Ver 1 1506...

Page 2: ...h component list 4 Chassie with component list 5 Spare parts 6 9 Spare part list 6 7 Replacing spare parts 8 9 Periodic inspection 10 14 Detail descriptions for Periodic inspection 10 Periodic inspection instruction 11 14 Care of the product 15 Technical specifications 15 2 SystemRoMedic TM ...

Page 3: ... wing handles are securely tightened before using the ReTurn General Before the product is used check to ensure that the screws are securely tightened The product must be inspected regularly by an authorized technician The product must be inspected thoroughly at least once each year or more regularly if it is used very frequently Inspection and service must be performed by an authorized technician...

Page 4: ...M6x16 6 Label SWL 7 Leg support Components 4 SystemRoMedic TM pos Title Art nr 1 Raising ladder 90001117 SHEET 1 OF 1 2 3 4 3 2 5 C D 4 6 7 8 A CHK D RMD0A0 F_Complete 23 09 2010 B WEIGHT I Atanasov A3 SCALE 1 10 DWG NO TITLE REVISION DO NOT SCALE DRAWING MATERIAL DATE SIGNATURE NAME APPV D MFG EDGES FINISH BREAK SHARP DEBUR AND Q A ANGULAR UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN MILLIMETERS ...

Page 5: ...2 mm vit POM 10 Center castors Tente UAP 125 mm 11 Screw Din 24015 8 8 M8x80s22 fzb 12 Nut MHM Din 1587 M8 fzb 13 Washer Ø16 Ø8x1 5 mm 14 Washer Ø20 Ø10x9 mm svart POM 15 Screw MVBF Din 603 M8x30 fzb 16 Screw K6S ISO 7380 M6x20 17 Nut DIN985 M6 fzb 18 Brake shoe antislip 19 Brake pad green 20 Brake pad red 21 Instruction label 22 Label ReTurn7600 Exploded view Chassie Components 5 SystemRoMedic TM...

Page 6: ...astor swirvel 75mm 1 set 4 x Castor swirvel 75mm 4 x Screw 4 x Locking nut 4 x Washer 7326 Locking pins 1 set 10 x Locking pin for leg support 7327 Instruction label 1 set 10 x Instruction label 7328 Brake shoe antislip 0 9 mm 1 set 10 x Brake shoe antislip 0 9 7329 Brake shoe antislip 1 5 mm 1 set 10 x Brake shoe antislip 1 5 7330 Brake kit L and R 1 set 1 x Brake pedal left 1 x Brake pedal right...

Page 7: ... no Product Unit Components Images 7334 Carriage bolt washers 1 set 2 x Carriage bolt M8x35 2 x Teflon coated washer 4 x Steel washer 2 x Locking acorn nut DIN 986 2 x Washer black 7334 x 2 x 2 x 2 x 4 x 2 7 SystemRoMedic TM ...

Page 8: ...rew the wing knobs and replace them with new ones If the stop plate needs to be replaced remove it and replace it with a new one Wing handles and stop for wing handles Unscrew the lower leg support from its mount Remove the lower leg support and mount and then install and secure the new ones The lower leg support is secured with 4 screws which are included in the spare parts kit Lower leg support ...

Page 9: ...neath The nut forms a unit with the locking pin so that when it is loosened the locking pin is also removed Secure the new locking pin and tighten the nut Locking pin For assembly instructions see page 12 under the Periodic Inspection section Antislip brake shoe For assembly instructions see page 13 under the Periodic Inspection section Brake kit For assembly instructions see page 12 under the Per...

Page 10: ... 3 Base plate Page 8 4 Brakes on fixed wheels Page 9 5 Fixed castors 125mm and axles 2 pcs Page 9 6 Castors 75mm 4 pcs Page 10 7 Leg support Page 10 Periodic inspection Detail descriptions for Periodic inspection 10 SystemRoMedic TM 3 6 5 4 7 2 1 ...

Page 11: ...en the ladder and undercarriage 2 pcs are securely tightened Base plate Turn the ReTurn around and check to ensure that all parts are securely in place The gap between the front and rear wheels should be between 2 3 mm Park the ReTurn on a level surface and place a 2 mm washer under one of the castor wheels The gap can be adjusted with the help of spacers on the castors Periodic inspection instruc...

Page 12: ...brake pedal fails to remain in the raised position even when the carriage bolt is adjusted the plastic washer can be replaced with a steel washer and a Teflon washer to increase tension on the brake pedal These are placed between the brake yoke and the chassis If the washers are replaced a longer carriage bolt must also be used See order of assembly of parts below Name of spare part Carriage bolt ...

Page 13: ...l for example a hex key to hold the steel washer and Teflon washer in place Ensure that the Teflon side is against the steel washer The steel washer must be turned so that the round side is against the Teflon washer 4 Install the pedal and hold the washers in place using the tool Insert the plastic bushing between the pedal s yoke and the chassis and then insert the tool to hold it in place 5 Inse...

Page 14: ...cs rotate and swivel easily If necessary remove dirt and hair from wheels Leg support Check both parallel and horizontal adjustment Friction can be adjusted by loosening or tightening the bolts at the rear of the support Castor swirvel Castor Castor Castor swirvel Castor swirvel Castor swirvel 14 SystemRoMedic TM ...

Page 15: ...oo to clean the ReTurn Do not use abrasive cloths or brushes to clean the base plate Do not use solvents To disinfect use 70 alcohol Technical specifications Art nr 7600 Complete ReTurn 7601 Ladder to 7600 SWL Max weight 200 kg 440 lbs static load Total weight 26 kg 57 lbs Patented 15 SystemRoMedic TM ...

Page 16: ...nd reduction of occupational injuries while allowing users to experience a greater sense of independence and dignity Through a unique combination of training and a complete range of efficient transfer aids SystemRoMedic offers improvement of both work environment and quality of care and at the same time achieves significant cost savings Always make sure that you have the correct version of the man...