Freeze-dryer Gamma 1-16 LSCplus
Freeze-dryer Gamma 2-16 LSCplus
6 Operation
54 / 126
Version 04/2013, Rev. 2.10 of 09/03/2020
• sb-dc
Translation of the original operating manual
Performance test (1)
The performance test is used to determine the following performance
parameters:
• vacuum decrease rate
• final vacuum
• minimum ice condenser temperature
NOTE
Prior to performing a test, ensure that the chamber is dry and unloaded
and that the ice condenser is defrosted.
Procedure:
• In the main window "Manual", select the button "Operating mode:
select/start"
– "Standard unit test".
• Select "Performance test" and start the test via the "Run" button.
The test will be performed. The parameters will be measured at defined
points of time, evaluated, and displayed in a dialog box (see the following
illustration).
If the freeze-dryer is equipped with the LPCplus SCADA software, the
results will be documented in the event list.
Fig. 30: Results of the performance tests
Evaluation:
Please contact the Martin Christ Gefriertrocknungsanlagen GmbH for an
assessment of the results.