Freeze-dryer Gamma 1-16 LSCplus
Freeze-dryer Gamma 2-16 LSCplus
6 Operation
Version 04/2013, Rev. 2.10 of 09/03/2020
• sb-dc
51 / 126
Translation of the original operating manual
Dialog box "Process and equipment messages" (7)
This dialog box shows and saves all of the error messages and other
messages. In the
event of an error or message, the window “Process and
equipment messages” will open automatically. In addition, a sound signal is
emitted until the error is acknowledged.
Malfunctions are divided into three categories:
• Red:
error messages
• Orange: process messages
• Yellow: general messages
The representation of the message provides information on its current
status. A double frame around a message means that the error has not
been eliminated yet. The colour of the button "quit" changes from blue to
grey once the message has been acknowledged.
Fig. 25: Representation of an error message
The advantage of this system is that malfunctions that occurred during the
night can be discovered the next day even if the cause of the malfunction
has already been eliminated.
The dialog box cannot be quit until all of the messages have been
acknowledged.
If a message has been acknowledged although the malfunction has not
been eliminated, the button "Process and equipment messages" will be
displayed in the respective colour of the malfunction in the main window.
1
The malfunction
has been
eliminated, but the
message has not
been acknowledged
yet
2
The malfunction
has not be
eliminated and the
message has not
been acknowledged
yet
3
The malfunction
has not been
eliminated yet, but
the message has
been acknowledged
Fig. 26: Dialog box "Process and equipment messages"