CAUTION
No modification of this device is allowed.
CAUTION
The manufacturer is held responsible for the fundamental safety, reliability and
performance of the device only if:
- The electrical system of the premises complies with the appropriate regulations;
- The device is used in accordance with the instructions for use.
Installation of the laser therapy equipment is fast and simple .
If the client requires an external safety interlock is necessary to expect a twisted pair
cabling of diameter 0.6mm minimum, 20m max, with screen connected to ground .
At the side safety circuit it necessary to plan the creation of a micro-switch to a via
normally closed .
Such circuit is an external safety accessory: it allows to interrupt the laser therapy
treatments if the door of the treatment room has been opened .
If the treatment room has only one door, the referring diagram is the following one :
Fig. D
If the treatment room has more than one door, the referring diagram is the following
one :
Fig. E
! WARNING !
If you want to install an extern interlock circuit ,
contact exclusively qualified technicians
and supply them the scheme correspondent to
the room used for the emission of the treatment.
A bad installation of the device can to generate
serious ocular lesions .
The following environmental conditions are ideal when installing the FISIOLASER
IRD-series unit :
-
room temperature
from +10° to +40°C
-
humidity level
from 10% to 80% without
condensation
-
avoid direct exposure to sunlight, chemical products
and vibrations