11 / 39
FISIOLASER IRD/IRD 2
! WARNING !
In the indicated temperature range, during normal use, laser handpieces can reach
temperatures of 53 ° without endangering patient health.
CAUTION
Do not use FISIOLASER IRD-series equipment
wherever there is a risk of either the external
or internal parts of the unit becoming wet
CAUTION
In the event of a blockage due to the intelock connector, after reconnecting the
interlock in the appropriate connector, press the START key only after taking the
handpiece in hand and repositioning it on the application part.
ELECTRICAL CONNECTION
The power entry module can be found on the back of the unit and consists of a three-
pole socket for the cable set, an extractible fuse box with two fuses (see technical
specifications) and the main switch.
Insert the three-pole plug of the cable set into the power entry module correctly.
!WARNING!
Before plugging the cable set into the mains,
check to make sure the equipment has not undergone
any damage during transport and make sure the
voltage corresponds to that of the unit (see the
rating plate on the side of the unit).
!WARNING!
The power supply to the unit is
VERY DANGEROUS!!
Make sure that the cable has been unplugged
from the mains supply before connecting or disconnecting from the unit.
!WARNING!
For safety purposes the cable set is provided
with an earthed plug.
Make sure that power supply socket has been earthed.
The equipment must only be connected
to power supply units that comply
with all safety directives in force.
When using an extension lead, make sure
that it has been earthed
Failure to comply with the above instructions
may lead to dangerous electrical discharge
causing machine damage and harm to persons.
CAUTION
If an extension lead is used with
FISIOLASER IRD-series and other equipment,
make sure that the total ampere absorption of all the devices together does not
exceed that consented for that type of cable, and in any case does not exceed 15A.
Once you have checked that installation and assembly have been carried out
according to instructions provided up to this point in the manual, switch on the
machine making sure that the display screen is turned on correctly .
SAFETY KEY AND SMART-CARD
The unit is supplied with an
(INTERLOCK)
safety key
comprising a DIN plug to
plug into the DIN socket on the back (Fig.1).
Fig. 1 :
Interlock