Appendix D Electro-magnetic compatibility tables
Guidance and manufacturer’s declaration – electromagnetic emissions
FOR ALL ME EQUIPMENT
The FISIOLASER IRD-series ME EQUIPMENT is intended for use in the electromagnetic
environment specified below.
The customer or the user of the FISIOLASER IRD-series ME EQUIPMENT should assure that it
is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF Emissions
CISPR 11
Group 1
The
FISIOLASER
IRD-series
ME
EQUIPMENT uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B
The
FISIOLASER
IRD-series
ME
EQUIPMENT is suitable for use in all
establishments,
including
domestic
establishments and those directly connected to
the public low voltage power supply network
that supplies buildings used for domestic
purposes .
Harmonic emissions
IEC 61000-3-2
Class A
The
FISIOLASER
IRD-series
ME
EQUIPMENT is suitable for use in all
establishments,
including
domestic
establishments and those directly connected to
the public low voltage power supply network
that supplies buildings used for domestic
purposes .
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
The
FISIOLASER
IRD-series
ME
EQUIPMENT is suitable for use in all
establishments,
including
domestic
establishments and those directly connected to
the public low voltage power supply network
that supplies buildings used for domestic
purposes.
13.4.1 Guidance and manufacturer’s declaration – electromagnetic immunity
FOR ALL ME EQUIPMENT
The FISIOLASER IRD-series ME EQUIPMENT is intended for use in the electromagnetic
environment specified below.
The customer or the user of the FISIOLASER IRD-series ME EQUIPMENT should assure that
it is used in such an environment.
Immunity test
IEC 60601
Test level
Compliance
level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6kV contact
Complies
Floors should be wood, concrete
or ceramic tile.
If floors are covered with
synthetic material , the relative
humidity should be at least 30 %.
± 8kV air
Complies
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV per power
supply lines
Complies
Mains power quality should be
that of a typical commercial or
hospital environment
± 1kV for
input / output lines
Complies
Surge
IEC 61000-4-5
± 1kV line(s) to
line(s)
Complies
Mains power quality should be
that of a typical commercial or
hospital environment
± 2kV line(s) to
earth
Complies
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% UT (>95% dip
in UT ) for 0.5
cycles
Complies
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the FISIOLASER IRD-series
ME
EQUIPMENT
requires
continued operation during power
mains
interruptions,
it
is
recommended
that
the
FISIOLASER IRD-series ME
EQUIPMENT be powered from
an uninterruptible power supply
or a battery.
40% UT (60% dip
in UT ) for 5
cycles
Complies
70% UT (30% dip
in UT ) for 25
cycles
Complies
<5% UT (>95% dip
in UT ) for 5 sec
Complies
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A / m
Complies
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial
or
hospital
environment
NOTE : UT is the a.c. mains voltage prior to application of the test level.