31 / 39
FISIOLASER IRD/IRD 2
CAUTION
In order to make a correct substitution of the fuses housed in FISIOLASER IRD-
series unit , please follow these instructions :
-
disconnect the power supply and open the fuse box using a screwdriver,
makingsure you insert the screwdriver in the slot on the fuse box and
levering up outwards;
-
insert a screwdriver into the two side holes for fuse expulsion
-
remove the old fuses
-
insert a new fuse at a time by using a slight pressure to the left, with a
finger
-
push the box back to fit into the slot.
! WARNING !
We recommend that the fuses substitution
is performed by qualified technical personnel ,
in order to execute this operation in safety conditions .
CAUTION
It is advisable to carry out periodic maintenance (every two years), making sure:
the intensity of any leakage currents
the continuity / integrity ground wire
the correctness of the value of insulation resistance
the characteristics of the laser emission
perform such audits ensures electrical safety of FISIOLASER IRD-series unit ,
ensuring that the unit operates under security conditions guaranteed, and allows you
to verify the correctness of the calibration of laser diodes .
For this kind of interventions we suggest to contact a technical and qualified service,
or in alternative contact CHINESPORT SpA or its authorized service centre .
! WARNING !
With the purpose to guarantee the operation of
the device under conditions of absolute safety
for the patient, we recommend to submit the device
to a cycle of periodic verifications (lilt at least 2 years),
because the device contains some parts submitted
to an electric degrade or aging, as the laser source if is constantly used under
conditions by maximum power .
The expected work life of device is 10 years.
TECHNICAL PROBLEMS
FISIOLASER IRD-series laser-therapy equipment has been designed and
manufactured using highly advanced technology and first class components for
reliable and efficient performance.
However, should you meet with any operational problems, we recommended that
you consult the following guide before contacting any of our authorised service
centres .
! WARNING !
DO NOT OPEN
FISIOLASER IRD-series equipment as
HIGH VOLTAGE ELECTRICITY is present
and may prove VERY DANGEROUS.
! WARNING !
DO NOT OPEN the handpiece/applicator :
this will cause damage to the accessories
and cause the warranty to lapse
CAUTION
Only CHINESPORT SpA authorised technicians may carry out service on the
internal parts of the equipment.
! WARNING !
The equipment contains an IR LASER SOURCE
which emits dangerous radiation .
For repairs and further information, contact CHINESPORT SpA or authorised
service centres .