Particular reference conditions:
The (+COM) input must not have been overloaded following the accidental application of a voltage to the input terminals with
the switch set
to Ω or T°. If this happens, the return to normal may take about ten minutes.
Protection : 1414 Vpk
Range
Accuracy
Resolution Protection
1000
0,1 % R + 8 D
10 m
1414 Vpk
10 k
0,07 % R + 8 D
100 m
100 k
1
1000 k
10
10 M
1 %
R + 80 D
100
100 M
3 %
R
+ 80 D
R ≤ 50 M
1 k
Automatic or manual range selection
“Active” protection by PTC thermistor
Measurement voltage: approx. 1.2V
Maximum open-circuit voltage: 3.5V typ.
In the 100M
range, in order to avoid the influence of the mains and guarantee the stated specifications, it is best to
disconnect the multimeter from the Wall Plug.
8.6.2.
100
Ω measure
Range
Accuracy
Resolution Protection
100
0,2 % R + 10 D
0,01
1414 Vpk
Capacity
8.7.
8.7.1.
Capacitance meter
In this setting, the user can measure the capacitance of a capacitor.
Range Operating ranget Specified measurement range Resolution
Intrinsic error
Measurement current Measurement time
1 nF
0 à 1,000 nF
0,100 à 1,000 nF
1 pF
2,5 % R
±
15 D
< 10 µA
≈ 400 ms
10 nF
0 à 10 nF
0,1 à 10,00 nF
10 pF
1 % R
±
8 D
< 10 µA
≈ 400 ms
100 nF
0 à 100,0 nF
1 à 100,0 nF
0,1 nF 1 % R
±
8 D
< 50 µA
≈ 400 ms
1000nF
0 à 1000nF
10 à 1000nF
1 nF
1 % R
±
10 D
< 200 µA
≈ 0,125 s/µF
10 µF
0 à 10,00 µF
1 à 10,00 µF
0,01 µF 1 % R
±
10 D
< 200 µA
≈ 0,125 s/µF
100 µF
0 à 100,0 µF
1 à 100,0 µF
0,1 µF
1 % R
±
10 D
< 500 µA
≈ 0,125 s/µF
1 mF
0 à 1,000 mF
0,1 à 1,000 mF
1 µF
1 % R
±
15 D
< 500 µA
≈ 17 s/mF
10 mF
0 à 10,00 mF
0,5 à 10,00 mF
10 µF
1,5 % R
± 1
5 D
< 500 µA
≈ 17 s/mF
(
) Use the REL function for values <10% of the range in order to restore the residual zero (compensation for the
capacitance of the leads)
Resolution 1,000 points
Automatic or manual range selection
“Active” protection by PTC thermistor
Maximum open-circuit voltage: 1V typ./4V max.