TECHNICAL CHARACTERISTICS
DC voltage
8.1.
In the “DC” mode, you measure a direct voltage or the DC component of an AC voltage.
The 100mV range is available only in manual mode, by
8.1.1.
MTX 3292B
Range
Input impedence Resolution Protection
Accuracy
100 mV (
) 10 M
/ 1 G
1 µV
1414 Vpk
0,1 % R + 30 D
1000 mV
11 M
/
1 G
10 µV
0,05 % R + 8 D
10 V
10,5 M
0,1 mV
0,03 % R + 8 D
100 V
10 M
1,0 mV
1000 V
10 M
10 mV
0,035 % R + 8 D
(
) REL mode activated (
measurement)
-
Recovery after triggering of the protection (>10V) approx. 10 s.
-
Protection 1 minute max.
-
Specifications valid from 0% to 100% of the range
-
Rejection : 100mV range common mode: > 40dB at 50Hz and 60Hz
-
1V range common mode: > 70dB at 50Hz and 60Hz
-
10V range common mode: >100dB at 50Hz and 60Hz
-
serial mode: > 60dB at 50Hz and 60Hz
-
Automatic or manual selection of the ranges
-
Protection by varistors
8.1.2.
MTX 3293B
Range
Input impedence Resolution Protection
Accuracy
100 mV (
) 10 M
/ 1 G
1 µV
1414 Vpk
0,1 % R + 30 D
1000 mV
10 M
/
1 G
10 µV
0,05 % R + 8 D
10 V
10,5 M
0,1 mV
0,02 % R + 8 D
100 V
10 M
1,0 mV
1000 V
10 M
10 mV
0,03 % R + 8 D
(
) REL mode activated (
measurement)
-
Recovery after triggering of the protection (>10V) approx. 10 s.
-
Protection 1 minute max.
-
Specifications valid from 0% to 100% of the range
-
Rejection: 100mV range common mode: > 40dB at 50Hz and 60Hz
-
1V range common mode: > 70dB at 50Hz and 60Hz
-
10V range common mode: >100dB at 50Hz and 60Hz