8.2.2.
MTX 3293B
Range Input impedence Resolution
Accuracy
45 Hz
to 1 kHz
1 to 100 kHz
100 to
200 kHz
100 mV (
)
10 M
Ω
1
μV
1 % R ± 50D 1 % R + 0,05 % x [F(kHz) - 1] R ± 50D (
)
-
1000 mV
11 M
Ω
10
μV
0,5 % R ± 40D
0,5%R + 0,2% x [F(kHz) - 1]R ± 40D
<10kHz
2,3%R + 0,02% x [F(kHz) - 10]R ± 40D
>10kHz
12 % R ± 50 D
(
)
10 V
10,5 M
Ω
0,1 mV
0,3 % R ± 30D 0,3 % R + 0,03 % x [F(kHz) - 1] R ± 30D 10 % R ± 30D
100 V
10 M
Ω
1 mV
0,3 % R ± 30D 0,3 % R + 0,015 % x [F(kHz) - 1] R ± 30D 8 % R ± 30D
1000 V (
)
10 M
Ω
10 mV
0,3 % R ± 30D 0,3 % R + 0,01 % x [F(kHz) - 1] R ± 30D
-
(
)
limitation at high frequency
(
) not contractual indicative values (see curves below)
(
)BP: Freq [kHz] limited to: 15,000/U input [V]
U input [V] limited to: 15,000/Freq [kHz]
Example: U input = 1,000 V
AC
Max. frequency: 15,000/1,000 = 15kHz
In the presence of a DC component: Additional error: (UDC/U measured) x (0.7% L + 70 D)
Example: UDC = 2V, U measured = 5 Vrms
Additional error: 0.28% L + 28 D
-
Rejection: common mode >80dB at 50Hz or 60Hz depending on selection
-
Automatic or manual selection of the ranges
-
Protection by varistors
-
Maximum acceptable permanent voltage: 1,414 Vpk
-
Specifications valid from: 10 to 100% of the range in the band from 20kHz to 200kHz
-
Influence of the peak factor on the accuracy in V
AC
, V at 50% of the range:
-
1% for a peak factor < 3.
As soon as the PEAK symbol appears, use the AUTO PEAK mode.
DC current
8.3.
Three possible modes: DC, AC, AC+DC
In DC mode, you can measure a direct current or the DC component of an alternating current.
In the AC and AC+DC modes, you can measure the true RMS (TRMS) value of an alternating current with/without its direct
component (no capacitive coupling in "DC" mode).
Fuse: ex.: SIBA/5019906/11A (10x38-11000-DMI-30kA-CR 1000V, very rapid action).
Range Input impedence Resolution Protection
Accuracy
1000 µA
170
10 nA
11 A
20 A < 30 s
0,1 % R + 15 D
10 mA
17
0,1 µA
0,08 %
R + 8 D
100 mA
1,7
1 µA
1000 mA
0,17
10 µA
0,15 % R + 8 D
10 A
0,03
(
)
100 µA
0,5 %
R + 15 D
100 A (
)
1000 µA
Range 100mV AC+DC
-6
-5
-4
-3
-2
-1
0
1
0
20
40
60
80
100
120
kHz
%
10mV
80mV
Range 1V AC+DC
-2
-1,5
-1
-0,5
0
0,5
1
1,5
2
0
20
40
60
80
100
120
kHz
%
100mV
800mV