44
Range
Resolution
Protection
Accuracy
10 à 100 Hz
0,001 Hz
1414 Vpk
0,02 % ± 10 D
100 à 1000 Hz
0,01 Hz
1000 Hz à 10 kHz
0,1 Hz
10 à 100 kHz
1 Hz
100 à 1000 kHz
10 Hz
1 MHz à 5 MHz
100 Hz
Range
Sensitivity (applicable only to rectangular signals)
100 mV
1 V
10 V
100 V
1000 V
0 Hz à 10 Hz
-
-
-
-
-
10 Hz à 200 kHz
10 %
20 à 5%
5 %
5 %
5 % (
)
200 à 500 kHz
20 %
5 %
5 à 2 % 5 à 10 % (
) 5 % (
)
500 à 1000 kHz
-
5%
2 %
10 %
5 % (
)
1 MHz à 5 MHz
2 à 50 %
20 % (
)
(
) Freq [kHz] limited to: 15,000/U input [V]
U input [V] limited to:
15,000/Freq [kHz]
(
) limited to 200kHz
The measurement is made by capacitive coupling.
Manual selection of freq. range, F <200kHz (default) or F >200kHz by a short press.
Input resistance:
10M
(Freq <100Hz)
Max. acceptable permanent voltage: 1,414 Vpk, see (
)
Protection by varistors on the voltage input.
8.5.2.
Secondary frequency measurement
Range
Resolution
Accuracy
Acceptable overload
10 à 100 Hz
0,001 Hz
0,02 % + 8 D
1450 Vcc
(1 min max.)
in range 100 mV
100 à 1000 Hz
0,01 Hz
1000 à 10 kHz
0,1 Hz
10 à 100 kHz
1 Hz
100 à 200 kHz
10 Hz
Range
Sensitivity (applicable only to rectangular signals) Vrms
100 mV
1 V
10 V à 1000 V (
)
1000 µA à 20 A (
)
10 Hz à 200 kHz
15% of the calibration 10% of the calibration 10% of the calibration
5 à 10 %
10 Hz à 10 kHz
10 kHz à 30 kHz
(
) Freq limited to [kHz]: 15,000/U input [V]
U input [V] limited to [V]: 15,000/Freq [kHz]
(
) at 50kHz for the «Ampere» range
The measurement is made by capacitive coupling.
Input resistance:
10M
(F <100Hz)
Protection by varistors on the voltage input.
Resistance
8.6.
8.6.1.
Ohmmeter
In this setting, the user can measure a resistance.