5
EN
CAUTION
• This unit should be operated at +5°C to +40°C and 15% to 90% Relative Humidity� The unit should be
transported and stored at -20°C to +60°C and 10% to 90% Relative Humidity�
• Use of parts or materials other than DJO’s can degrade minimum safety�
• Connect to this unit only items and equipment that have been specified in this IFU as part of the ME
SYSTEM or that have been specified as being compatible with the ME SYSTEM�
• DO NOT disassemble, modify, or remodel the unit or accessories� This may cause unit damage, malfunction,
electrical shock, fire, or personal injury�
• DO NOT permit foreign materials, liquids or cleaning agents to enter the unit, including, but not limited to,
inflammables, water, and metallic objects, to prevent unit damage, malfunction, electrical shock, fire, or
personal injury�
• Before each use, inspect Ultrasound Applicator for cracks, which may allow the ingress of conductive fluid�
• Before each use, inspect Applicator cables, STIM cables and associated connectors�
• Before each use, inspect Vacuum Electrode Cups and Lead Hoses for cracks and damage which may not
allow the vacuum to properly secure the electrodes�
• Handle Ultrasound Applicator with care� Inappropriate handling may adversely affect its characteristics�
• Caution should always be exercised with current densities more than 2mA/cm²�
• There are no user-serviceable parts inside the unit� If a malfunction occurs, discontinue use immediately
and consult dealer for repair service�
• In case of device unused with battery embedded, it is recommended to connect the device at least once
every 4 months to allow battery recharge�
• For waveforms with a DC component:
»
Do not shave electrodes application area
»
Warn the patient that tingling sensation under electrodes is normal and it is not linked to burn risk�
»
Rinse thoroughly treatment area with tap water immediately after the treatment
Summary of Contents for Intelect Mobile 2 Ultrasound
Page 31: ...31 EN COMBO STIM ULTRASOUND Guide de démarrage rapide FR ...
Page 61: ...61 FR COMBO STIM ULTRASOUND Guía de inicio rápido ES ...
Page 91: ...91 ES COMBO STIM ULTRASOUND Guida rapida d avvio IT ...
Page 121: ...121 IT COMBO STIM ULTRASOUND Kurzanleitung DE ...
Page 151: ...143 DE COMBO STIM ULTRASOUND Snelle start gids NL ...
Page 181: ...173 NL COMBO STIM ULTRASOUND Hızlı başlama rehberi TR ...
Page 211: ...203 TR COMBO STIM ULTRASOUND Snabbstartsguide SE ...
Page 241: ...233 SE COMBO STIM ULTRASOUND Pikaopas FI ...
Page 271: ...263 FI COMBO STIM ULTRASOUND Hurtig start guide DA ...
Page 301: ...293 DA COMBO STIM ULTRASOUND Guia rápido PT ...
Page 331: ...323 PT COMBO STIM ULTRASOUND Rychlý návod CZ ...
Page 361: ...353 CZ COMBO STIM ULTRASOUND Skrócona instrukcja obsługi PO ...
Page 391: ...383 PO COMBO STIM ULTRASOUND Οδηγός γρήγορης εκκίνησης GR ...
Page 421: ...413 GR COMBO STIM ULTRASOUND Краткое руководство пользователя RU ...
Page 451: ...443 RU COMBO STIM ULTRASOUND Ghid de inițiere rapidă RO ...
Page 481: ...473 RO ...