27
EN
WAVEFORMS (CONTINUED)
MONOPHASIC: Monophasic Rectangular
Pulsed
Advice on size and type of electrodes to be used is given in device GUI
"treatment guidelines" feature
The Monophasic Rectangular Pulsed waveform is an interrupted
unidirectional current with a rectangular pulse shape.
Output Mode
Electrodes
Available on Channels
1, 2
Treatment Time
1-60 min
Mode Selection
CC
Output Intensity
0-60 mA (CC)
Phase Duration
0.1-1,000 ms
Phase Interval
5-5,000 ms
IRMS 0-47mA
DC component
Yes
GALVANIC: Continuous
Advice on size and type of electrodes to be used is given in device GUI
"treatment guidelines" feature
Galvanic Current is a direct current flowing in one direction only.
The current can be continuous or interrupted.
Output Mode
Electrodes
Available on Channels
1, 2
Treatment Time
1-60 min
Mode Selection
CC
Output Intensity
0-40 mA (CC)
Cycle Time
Continuous, or User Defined
Polarity Reversal
On or Off
With Polarity Reversal On, Polarity will change in
the middle of the treatment time.
IRMS 0-44mA
DC component
Yes
MONOPHASIC: Monophasic Triangular
Pulsed
Advice on size and type of electrodes to be used is given in device GUI
"treatment guidelines" feature
The Monophasic Triangular Pulsed waveform is an interrupted
unidirectional current with a triangular pulse shape.
Output Mode
Electrodes
Available on Channels
1, 2
Treatment Time
1-60 min
Mode Selection
CC
Output Intensity
0-60 mA (CC)
Phase Duration
0.1-1,000 ms
Phase Interval
5-5,000 ms
IRMS 0-27mA
DC component
Yes
GALVANIC: Interrupted
Advice on size and type of electrodes to be used is given in device GUI
"treatment guidelines" feature
Galvanic Current is a direct current flowing in one direction only.
The current can be continuous or interrupted.
Output Mode
Electrodes
Available on Channels
1, 2
Treatment Time
1-60 min
Mode Selection
CC
Output Intensity
0-40 mA (CC)
Pulse Duration
136 μsec
Phase Interval
25 usec
Polarity Reversal
On or Off
With Polarity Reversal On, Polarity will
change in the middle of the treatment time.
Polarity Reversal Ramp
1 sec
IRMS 0-41mA
DC component
Yes
Summary of Contents for Intelect Mobile 2 Ultrasound
Page 31: ...31 EN COMBO STIM ULTRASOUND Guide de démarrage rapide FR ...
Page 61: ...61 FR COMBO STIM ULTRASOUND Guía de inicio rápido ES ...
Page 91: ...91 ES COMBO STIM ULTRASOUND Guida rapida d avvio IT ...
Page 121: ...121 IT COMBO STIM ULTRASOUND Kurzanleitung DE ...
Page 151: ...143 DE COMBO STIM ULTRASOUND Snelle start gids NL ...
Page 181: ...173 NL COMBO STIM ULTRASOUND Hızlı başlama rehberi TR ...
Page 211: ...203 TR COMBO STIM ULTRASOUND Snabbstartsguide SE ...
Page 241: ...233 SE COMBO STIM ULTRASOUND Pikaopas FI ...
Page 271: ...263 FI COMBO STIM ULTRASOUND Hurtig start guide DA ...
Page 301: ...293 DA COMBO STIM ULTRASOUND Guia rápido PT ...
Page 331: ...323 PT COMBO STIM ULTRASOUND Rychlý návod CZ ...
Page 361: ...353 CZ COMBO STIM ULTRASOUND Skrócona instrukcja obsługi PO ...
Page 391: ...383 PO COMBO STIM ULTRASOUND Οδηγός γρήγορης εκκίνησης GR ...
Page 421: ...413 GR COMBO STIM ULTRASOUND Краткое руководство пользователя RU ...
Page 451: ...443 RU COMBO STIM ULTRASOUND Ghid de inițiere rapidă RO ...
Page 481: ...473 RO ...