30
INTELECT® MOBILE 2 QUICK START GUIDE
WAVEFORMS (CONTINUED)
Isoplanar Vector Interferential Current
Interferential Current is a medium frequency waveform. Current is
distributed through two channels (four electrodes). The currents
cross each other in the body at the area requiring treatment. The
two currents interfere with each other at this crossing point,
resulting in a modulation of the intensity (the current intensity
increases and decreases at a regular frequency). In Isoplanar
Vector IFC Channel B has a fixed phase shift of 45° against Channel
A.
Output Mode
Electrodes
Available on Channel
1&2
Treatment Time
1-60 Minutes
Amplitude
0-100 mA (CC)
Beat Frequency
1-200 Hz
Carrier Frequency
2000-10,000 Hz
Cycle Time
Continuous or User Defined
Ramp
0-5 s
Sweep Time
14 sec
Sweep Low Beat Frequency
1-199 Hz
Sweep High Beat Frequency
2-200 Hz
Vector Scan
Fixed at 45°
IRMS 0-55mA
DC component
No
Dipole Vector Interferential Current
Interferential Current is a medium frequency waveform. Current is
distributed through two channels (four electrodes). The currents
cross each other in the body at the area requiring treatment. The
two currents interfere with each other at this crossing point,
resulting in a modulation of the intensity (the current intensity
increases and decreases at a regular frequency). With the dipole
vector technique the currents from the two electrode pairs are
vectorially summed in the tissue. The effect is that stimulation
only occurs into the direction of the resulting vector, which can be
adjusted over a range of 360º.
The angle is either manually adjusted and constant, or
automatically generated so that the full revolution (360°) is
automatically generated in an adjustable timing (rotation time).
Output Mode
Electrodes
Available on Channel
1&2
Treatment Time
1-60 Minutes
Amplitude
0-100 mA (CC)
Beat Frequency
1-200 Hz
Carrier Frequency
2000-10,000 Hz
Cycle Time
Continuous or User Defined
Ramp
0-5 s
Resolution Vector angle (manual mode)
359°
Rotation Time (automatic mode)
1-10 s
Sweep Time
14 sec
Sweep Low Beat Frequency
1-199 Hz
Sweep High Beat Frequency
2-200 Hz
Vector Scan
Manual / Automatic
IRMS 0-55mA
DC component
No
Summary of Contents for Intelect Mobile 2 Ultrasound
Page 31: ...31 EN COMBO STIM ULTRASOUND Guide de démarrage rapide FR ...
Page 61: ...61 FR COMBO STIM ULTRASOUND Guía de inicio rápido ES ...
Page 91: ...91 ES COMBO STIM ULTRASOUND Guida rapida d avvio IT ...
Page 121: ...121 IT COMBO STIM ULTRASOUND Kurzanleitung DE ...
Page 151: ...143 DE COMBO STIM ULTRASOUND Snelle start gids NL ...
Page 181: ...173 NL COMBO STIM ULTRASOUND Hızlı başlama rehberi TR ...
Page 211: ...203 TR COMBO STIM ULTRASOUND Snabbstartsguide SE ...
Page 241: ...233 SE COMBO STIM ULTRASOUND Pikaopas FI ...
Page 271: ...263 FI COMBO STIM ULTRASOUND Hurtig start guide DA ...
Page 301: ...293 DA COMBO STIM ULTRASOUND Guia rápido PT ...
Page 331: ...323 PT COMBO STIM ULTRASOUND Rychlý návod CZ ...
Page 361: ...353 CZ COMBO STIM ULTRASOUND Skrócona instrukcja obsługi PO ...
Page 391: ...383 PO COMBO STIM ULTRASOUND Οδηγός γρήγορης εκκίνησης GR ...
Page 421: ...413 GR COMBO STIM ULTRASOUND Краткое руководство пользователя RU ...
Page 451: ...443 RU COMBO STIM ULTRASOUND Ghid de inițiere rapidă RO ...
Page 481: ...473 RO ...