3 SCPI
3.3 Instrument-specific Commands
54
Parameter description:
<time> Time, the default unit is s
Range: [0ms, 100s]
Example:
:SENSe:CONFigure:LOSCillator:PARameter:SETTling 50ms
:SENSe:CONFigure:LOSCillator:PARameter:SETTling?
Reset status:
100ms
Key path:
Front panel
【
Mode Setup
】
—>[Ext LO Config]—>[Settling Time]
[ External Local Oscillator Multiplier Value ]
:SENSe:CONFigure:LOSCillator:PARameter:MULTiplier
Function description:
This command is used to set/query the multiplier value of the external local
oscillator.
Setting format:
:SENSe:CONFigure:LOSCillator:PARameter:MULTiplier <integer>
Query format:
:SENSe:CONFigure:LOSCillator:PARameter:MULTiplier?
Return value:
Multiplier value
Parameter description:
<integer> Integer
Example:
:SENSe:CONFigure:LOSCillator:PARameter:MULTiplier 2
:SENSe:CONFigure:LOSCillator:PARameter:MULTiplier?
Reset status:
1
Key path:
Front panel
【
Mode Setup
】
—>[Ext LO Config]—>[Multiplier]
[ External Local Oscillator Divider Value ]
:SENSe:CONFigure:LOSCillator:PARameter:DIVider
Function description:
This command is used to set/query the divider value of the external local
oscillator.
Setting format:
:SENSe:CONFigure:LOSCillator:PARameter: DIVider <integer>
Query format:
:SENSe:CONFigure:LOSCillator:PARameter: DIVider?
Return value:
Divider value
Parameter description:
<integer> Integer
Example:
:SENSe:CONFigure:LOSCillator:PARameter:DIVider 2
:SENSe:CONFigure:LOSCillator:PARameter:DIVider?
Reset status:
1
Key path:
Front panel
【
Mode Setup
】
—>[Ext LO Config]—>[Divider]
[ External Local Oscillator Frequency Offset Value ]
:SENSe:CONFigure:LOSCillator:PARameter:OFFSet
Function description:
This command is used to set/query the frequency offset value of the external