3 SCPI
3.3 Instrument-specific Commands
53
3.3.5.2 External Local Oscillator Configuration
The external local oscillator configuration corresponds to the
【
Mode Setup
】
[Ext LO Config] key on the
front panel. It is used to set the related properties of the external local oscillator, including the external
local oscillator GPIB address, settling time, multiplier value, divider value and external local oscillator
frequency offset.
[ External Local Oscillator On-off Status ]
:SENSe:CONFigure:MODE:DUT:LO:CONTrol:STATe
Function description:
This command is used to set/query the on-off status of the external local
oscillator source.
Setting format:
:SENSe:CONFigure:MODE:DUT:LO:CONTrol:STATe OFF|ON|0|1
Query format:
:SENSe:CONFigure:MODE:DUT:LO:CONTrol:STATe?
Return value:
RF|IF
Parameter description:
OFF|0: Off
ON|1: On
Example:
:SENSe:CONFigure:MODE:DUT:LO:CONTrol:STATe OFF
:SENSe:CONFigure:MODE:DUT:LO:CONTrol:STATe?
Reset status:
OFF
Key path:
Front panel
【
Mode Setup
】
—>[DUT Setup]—>[Ext LO Config]
[ External Local Oscillator Input Power ]
:SENSe:CONFigure:MODE:DUT:EXTend:LO:POWer
Function description:
This command is used to set/query the input power of the external local oscillator.
The default unit is dBm.
Setting format:
:SENSe:CONFigure:MODE:DUT:EXTend:LO:POWer <value>
Query format:
:SENSe:CONFigure:MODE:DUT:EXTend:LO:POWer?
Return value:
Power value, the default unit is dBm
Parameter description:
<value> Power value.
Example:
:SENSe:CONFigure:MODE:DUT:EXTend:LO:POWer 0dBm
:SENSe:CONFigure:MODE:DUT:EXTend:LO:POWer?
Reset status:
-10.00 dBm
Key path:
Front panel
【
Mode Setup
】
—>[DUT Setup]—>[External LO Power]
[ External Local Oscillator Settling Time ]
:SENSe:CONFigure:LOSCillator:PARameter:SETTling[:TIME]
Function description:
This command is used to manage the settling time of the external local oscillator.
The unit may be ms or s.
Setting format:
:SENSe:CONFigure:LOSCillator:PARameter:SETTling[:TIME] <time>
Query format:
:SENSe:CONFigure:LOSCillator:PARameter:SETTling[:TIME]?
Return value:
The default unit is ms. If it is 100 ms, 0 will be returned