78
x
Press the
[POWER]
key in the peak detection standby state to initiate the power
standby state.
x
Press the
[POWER]
key during detection of the peak to return to the peak detection
standby state.
x
Peak Hold setting can be cleared without menu operation, by pressing the
[POWER]
key for about three seconds.
x
Polarity of the peak value displayed is “polarity of the displayed value of the first
change (by five times or more of Zero Range) from the display within Zero Range.”
x
Usually the peak value is easily measured by selecting menu item
23
(Pouring mode). Depending on the weighting conditions and the sample type,
this setting may not always be suitable.
x
Pretaring Value (13.7) function cannot be used with Peak Hold function.
x
See 13.5 for details of Zero Range.
x
See 11.3 for details of stability detection.
x
Not more than one of these seven functions can be set at a time. Setting the second
one with the menu will automatically clear the previously set function;
Auto Print(13.3), Auto Zero(13.4), Peak Hold(14.3), Interval Timer(14.4),
Add-on mode(14.5), Animal Weighing mode(14.6), Formulation mode(14.7).
x
When minimum display has been switched to 10d (See 9.2), the optimized
Zero Range for 1d display may cause failure in the operation.
Select optimal Zero Range value again when minimum display has been
changed (See 13.5 for details).
Summary of Contents for CU series
Page 1: ...2 ...
Page 12: ...13 c Small pan model minimum display 0 001g windbreak standard a b c common ...
Page 19: ...20 a Large pan model b Small pan model minimum display 0 01g ...
Page 20: ...21 c Small pan model minimum display 0 001g ...
Page 106: ...107 ...
Page 107: ...108 ...
Page 108: ...109 ...
Page 109: ...110 ...
Page 120: ...121 W Weighing 20 WindowsDirect 22 91 107 Z Zero range 60 Zero tracking 50 ...
Page 124: ...125 MEMO ...
Page 125: ...126 MEMO ...
Page 126: ...127 ...
Page 127: ...128 ...