13.2 Piece Counting (PCS)
1
. Set up the PCS with menu item
57
if it is not set.
(The PCS unit is set before shipment.)
2
. Press the
[UNIT]
key several times in the mass
display until the PCS is displayed.
3
. Load the container and press the
[O/T]
key to
tare the balance.
4
. Count exactly five pieces (or 10, 20, 50, 100, or
200 pieces) of sample to be measured and load
them on the pan.
(Example)
5
. Press the
[CAL]
key.
6
. Every time the
[CAL]
key is pressed, the display
sequentially changes as
“
Ld 5pcs” ....“Ld 200pcs”,
“Ld 5pcs”... The default setting is “Ld 10pcs”.
Press the
[O/T]
key when the display is equivalent
to the number of loaded pieces.
Example:
If 50 pieces are loaded, press the
[O/T]
key when “Ld 50pcs” is displayed. This
determines the unit weight or average
weight per piece.
As sample is added or removed, the
piece count (number of pieces) is displayed.
Repeat steps 3 through 6 above when the sample or manufacturing lot is changed.
In this example, when the
[CAL]
key is pressed in the next PCS menu, display
starts from “Ld 50pcs”.
Summary of Contents for CU series
Page 1: ...2 ...
Page 12: ...13 c Small pan model minimum display 0 001g windbreak standard a b c common ...
Page 19: ...20 a Large pan model b Small pan model minimum display 0 01g ...
Page 20: ...21 c Small pan model minimum display 0 001g ...
Page 106: ...107 ...
Page 107: ...108 ...
Page 108: ...109 ...
Page 109: ...110 ...
Page 120: ...121 W Weighing 20 WindowsDirect 22 91 107 Z Zero range 60 Zero tracking 50 ...
Page 124: ...125 MEMO ...
Page 125: ...126 MEMO ...
Page 126: ...127 ...
Page 127: ...128 ...