122
[POWER] key
[CAL ] key
[O/T] key
[UNIT] key
[PRINT] key
[Explanatory Operation Sheet]
Key Switch Basic Function
[POWER]
Switches between operation and stand-by (On/Off).
[CAL]
Enters span calibration or menu item selection.
[O/T]
Tares the balance (Displays zero with a container on the pan)
[UNIT]
Switches weighing units.
Output the displayed data to an electronic printer or a computer.
[PRINT]
To output date and time
→
Press and hold for about 3 seconds
Stabilizing Display / Faster Response for Filling
x
If it is unavoidable to install the balance at a location subject to air flow or vibration, try
changing the environmental settings to stabilize the reading.
(From mass display) Press [CAL] repeatedly till “E” is blinking in “SEL:EAUS” display
→
Press
[O/T] (“
→
E- AbtP8” is displayed with “
→
” blinking)
→
Press [CAL] once (“A” blinks)
→
Press
[CAL] repeatedly until “EA-vibr” is displayed
→
Press [O/T] (Stability mark is added to “EAvibr”)
→
Press [POWER] repeatedly to return to mass display.
This setting will bring more stability, however response will be slower. If further stability is
desired, select “EA-wind” instead of “EA-vibr” in the above procedure.
x
If faster response is required for filling or dosing purpose, select “PoUr” instead of “EA-vibr”
in the above procedure. Pouring mode is set and the response will be very fast, but display
unstable. Refer to 11.2 for details of Pouring mode.
Hints on Fully-automatic Span Calibration(UW series only)
Fully-automatic span calibration PSC
Default setting of PSC is ON. Upon detection of temperature change, span calibration is
automatically performed to maintain accuracy. The Weight symbol blinks prior to
fully-automatic calibration.
How to skip a PSC calibration
Press [POWER] key while Weight symbol is blinking.
Summary of Contents for CU series
Page 1: ...2 ...
Page 12: ...13 c Small pan model minimum display 0 001g windbreak standard a b c common ...
Page 19: ...20 a Large pan model b Small pan model minimum display 0 01g ...
Page 20: ...21 c Small pan model minimum display 0 001g ...
Page 106: ...107 ...
Page 107: ...108 ...
Page 108: ...109 ...
Page 109: ...110 ...
Page 120: ...121 W Weighing 20 WindowsDirect 22 91 107 Z Zero range 60 Zero tracking 50 ...
Page 124: ...125 MEMO ...
Page 125: ...126 MEMO ...
Page 126: ...127 ...
Page 127: ...128 ...