44
8. Built-in Clock Set-up
The built-in clock has to be set up in advance if a calibration record is to be produced or
Clock-CAL function is to be used.
8.1 Date
1
. Select menu item
63
and set the last two figures
of the year, month and day, using the
[UNIT]
and
[PRINT]
keys.
Example:May 15th, 2002, set as “02.05.15”
Example:February 29th, 2004, set as “04.02.29”
(Example)
2
. Then press the
[O/T]
key.
x
The built-in clock corrects for the leap year automatically.
x
The moment the [O/T] key is pressed to finish setting, seconds are set
to zero. If the is set after setting the time, the second value will be incorrect.
It is important to set the first and then the time, or to correct the seconds value using
the ± second correcting function described in section 8.3.
8.2 Date Output Style]
The order of the year, the month and the date in the external output can be selected from three styles.
The setting made here is not reflected on the display of the balance.
The setting made here is not reflected on the display of the balance.
To output in the YYYY-MM-DD order, select menu item
63a
. [y.m.d]
To output in the DD-MM-YYYY order, select menu item
63b
. [d.m.y]
To output in the MM-DD-YYYY order, select menu item
63c
. [m.d.y]
The setting made here on “Date Output Style” will not be cleared with Menu reset
(See 7.5.2).
Summary of Contents for CU series
Page 1: ...2 ...
Page 12: ...13 c Small pan model minimum display 0 001g windbreak standard a b c common ...
Page 19: ...20 a Large pan model b Small pan model minimum display 0 01g ...
Page 20: ...21 c Small pan model minimum display 0 001g ...
Page 106: ...107 ...
Page 107: ...108 ...
Page 108: ...109 ...
Page 109: ...110 ...
Page 120: ...121 W Weighing 20 WindowsDirect 22 91 107 Z Zero range 60 Zero tracking 50 ...
Page 124: ...125 MEMO ...
Page 125: ...126 MEMO ...
Page 126: ...127 ...
Page 127: ...128 ...