98
Restricted Items in the Multi-Connection Mode
x
Multi-Connection mode is not designed for each balance to independently send the data.
This mode is for sampling the data by control of multiple balances with one PC.
This is not the function to support multiple balances for sending the data separately.
When multiple number of data and commands exist on the system at the same time, the
balance may not operate normally. For example, multiple data may exist simultaneously
on the system when using multiple balances in the Application Measurement mode such
as Auto Print or Continuous Output, or pressing the
[PRINT]
key. Normal operation cannot
be guaranteed.
x
Peripheral instruments connected to the DATA I/O or IFB-102A connector, such as an
EP- 50 Electronic Printer, cannot be used.
x
The communication formats can only be used with “Format EB type”. Delimiters cannot
be used with menu items
97
or
98
(setting for WindowsDirect).
x
It will take an average 0.05×N seconds for the data from the personal computer to reach
balance unit N, even at setting of 38400 bps. Also, it will take almost the same time for the
data from the balance in the Nth position to reach the personal computer.
Cancellation of Multi-Connection mode
Multi-Connection mode cannot be canceled unless the power for all balances is turned OFF.
Summary of Contents for CU series
Page 1: ...2 ...
Page 12: ...13 c Small pan model minimum display 0 001g windbreak standard a b c common ...
Page 19: ...20 a Large pan model b Small pan model minimum display 0 01g ...
Page 20: ...21 c Small pan model minimum display 0 001g ...
Page 106: ...107 ...
Page 107: ...108 ...
Page 108: ...109 ...
Page 109: ...110 ...
Page 120: ...121 W Weighing 20 WindowsDirect 22 91 107 Z Zero range 60 Zero tracking 50 ...
Page 124: ...125 MEMO ...
Page 125: ...126 MEMO ...
Page 126: ...127 ...
Page 127: ...128 ...