14.5 Add-on Mode
This function is convenient for weighing a large number of sample components as they are
added on the balance. Every time a component is added and stability is detected, the mass of
it is displayed and output through RS-232C or DATA I/O interface. The display will then be
automatically zeroed for the next component. When all the components have been weighed,
the total mass is calculated by pressing
[POWER]
key.
Select menu item
52
.
1
. Load the weighing vessel and press the
[O/T]
key in
the Add-on standby state (The “Add-on” and
the Standby symbols are lit).
Zeroing occurs.
2
. Press the
[PRINT]
key.
The Stand-by symbol disappears, and Add-on
measurement starts.
3
. Load the first sample.
When stability is detected with a displayed value
five times the Zero Range or more, the displayed
value is automatically output and zeroing occurs.
Even if the displayed value is less than five times
the Zero Range, pressing
[PRINT]
key at stability
causes the same effect.
4
. Repeat the above step 3 until all the components
have been weighed.
5
. Press the
[POWER]
key.
The total sample mass on the pan is displayed and
the balance returns to the Add-on standby state.
Press the
[PRINT]
key to output the total mass.
Start the next set of measurements from the above
step
1
.
Summary of Contents for CU series
Page 1: ...2 ...
Page 12: ...13 c Small pan model minimum display 0 001g windbreak standard a b c common ...
Page 19: ...20 a Large pan model b Small pan model minimum display 0 01g ...
Page 20: ...21 c Small pan model minimum display 0 001g ...
Page 106: ...107 ...
Page 107: ...108 ...
Page 108: ...109 ...
Page 109: ...110 ...
Page 120: ...121 W Weighing 20 WindowsDirect 22 91 107 Z Zero range 60 Zero tracking 50 ...
Page 124: ...125 MEMO ...
Page 125: ...126 MEMO ...
Page 126: ...127 ...
Page 127: ...128 ...